Philips V60 Ventilators with Power Management PCBA part number 1055906

Publicado el 8 de abril de 2020 · Inicio del retiro: 5 de marzo de 2020

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Respironics California, LLC
Distribuidor
U.S.: CO, PA, AZ, OH, MT, AR, FL, NC, KS, IL, CA, HI, MD, SC, WY, OR, NY, NE, OK, MN, UT, ND, LA, IN, AL, AK, WV, GA, WI, MI, VA, NJ, TX, WA, SD, KY, TN, MS, MO, DE, MA, VT, CT, IA, NV, ID, ME, NM, DC, NH, VI, RI, PR. O.U.S.: Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Gibraltar, Hong Kong, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Jersey, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Kuwait, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Malta, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nigeria, Norway, Oman, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Viet Nam
Lote
Serial Numbers: 100002908 to 100027093; 100027096 to 100085389; 100085399 to 100109746; 100109868 to 201003390; 201003392 to 201010952
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States