Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1657-2013

***REF SC2100***eMax 2 Plus Console***SN G 10307653002*** Product Usage: The RPM of the Motor Handpiece may be less than 80,000 RPM (approximately 72,000 RPM) in the free running state when using the optional Hand Control. The low Motor Handpiece RPM would only be experienced in the Free running state (Hand Control fully depressed and no load on the cutting bur). This condition would not affect the power output of the Handpiece in the loaded working range (typically 50,000 to 70,000 RPM.

Publicado el 10 de julio de 2013 · Inicio del retiro: 24 de junio de 2011

Resumen

FDA registra el retiro Z-1657-2013 de dispositivos médicos de Clase II, publicado el 10 de julio de 2013 e iniciado el 24 de junio de 2011. Empresa responsable: The Anspach Effort, Inc.; Estado: Finalizado; Cantidad reportada: 61 units; Lote: SC2100 System Console Emax 2 Plus; SC2101 System Console, Single Port with irrigation; SC2102 System Console, Single Port, without Irrigation. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The RPM of the Motor Handpiece may be less than 80,000 RPM.

Qué hacer

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
The Anspach Effort, Inc.
Distribuidor
Worldwide Distribution - USA Nationwide including the states of: AZ, CA, FL, KY, ME, MI, MO, OH, PA, TN, TX , VA, and the countries of: Australia, Egypt, Italy, Japan, Switzerland, Taiwan, and Turkey;
Lote
SC2100 System Console Emax 2 Plus; SC2101 System Console, Single Port with irrigation; SC2102 System Console, Single Port, without Irrigation.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿***REF SC2100***eMax 2 Plus Console***SN G 10307653002*** Product Usage: The RPM of the Motor Handpiece may be less than 80,000 RPM (approximately 72,000 RPM) in the free running state when using the optional Hand Control. The low Motor Handpiece RPM would only be experienced in the Free running state (Hand Control fully depressed and no load on the cutting bur). This condition would not affect the power output of the Handpiece in the loaded working range (typically 50,000 to 70,000 RPM. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1657-2013 el 10 de julio de 2013. Peligro reportado: The RPM of the Motor Handpiece may be less than 80,000 RPM.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit