ARCHITECT Intact PTH Controls consists of 3 x 1 Bottle (8.0 mL each) of PTH (synthetic peptide) in Bis Tris Propane buffer with protein (bovine) stabilizer. Preservatives: sodium azide and ProClin 300. Product Usage: The ARCHITECT Intact PTH assay is an in vitro chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of intact parathyroid hormone (PTH) in human serum and plasma on the ARCHITECT i System.
Publicado el 4 de junio de 2014 · Inicio del retiro: 12 de febrero de 2014
Resumen
FDA registra el retiro Z-1672-2014 de dispositivos médicos de Clase II, publicado el 4 de junio de 2014 e iniciado el 12 de febrero de 2014. Empresa responsable: Abbott Laboratories; Estado: Finalizado; Cantidad reportada: List Number 8K25-10: 9,463 control kits; Lote: List Number: 8K25-10. Lot Numbers and Expiration Dates: Lot 00713I000, Expires 05 SEP 2014; Lot 01813C000, Expires 14 MAR 2014; Lot 01813K000, Expires 12 NOV 2014; Lot 02613E000, Expires 09 MAY 2014; Lot 01412K000, Expires 12 NOV 2013; Lot 00713A000, Expires 08 JAN 2014; Lot 01012F000, Expires 30 MAY 2013. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Abbott has confirmed that a performance shift in the ARCHITECT Intact PTH assay has the potential to generate falsely elevated results on patient samples. Results generated with impacted lots may demonstrate a positive shift relative to those generated with previous calibrator and/or control lots. The issue may also impact your established ARCHITECT Intact PTH reference ranges. Abbott ARCHITECT In
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Abbott Laboratories
- Distribuidor
- Worldwide Distribution - US Nationwide including District of Columbia and Puerto Rico; Algeria, Andorra, Angola, Argentina, Armenia, Austria, Azerbaijan, Bahamas, Bangladesh, Barbados, Belarus, Belgium, Bosnia & Herzegovina, Brazil, Brunei, Bulgaria, Cameroon, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Ethiopia, Finland, France, Gaza and Jericho, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, ITALY, Jamaica, Japan, Jordan, Kenya, Korea (South Korea), Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Mauritania, Mexico, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Panam®, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, San Marino, Saudi, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, SPAIN, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, Ukraine, Uruguay, Venezuela, Vietnam, and Yemen
- Lote
- List Number: 8K25-10. Lot Numbers and Expiration Dates: Lot 00713I000, Expires 05 SEP 2014; Lot 01813C000, Expires 14 MAR 2014; Lot 01813K000, Expires 12 NOV 2014; Lot 02613E000, Expires 09 MAY 2014; Lot 01412K000, Expires 12 NOV 2013; Lot 00713A000, Expires 08 JAN 2014; Lot 01012F000, Expires 30 MAY 2013.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAAlinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-2024 de diciembre de 2025
- FDABrand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-01 Software Version: All30 de abril de 2025
- FDAAlinity hq Analyzer, REF: 09P68-0128 de agosto de 2024
- FDAARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20.10 de abril de 2024
- FDAGREENSTONE BRAND fluconazole for Oral Suspension, ORANGE FLAVORED, 40 mg/mL, when reconstituted, Rx only,11 de junio de 2014
- FDAARCHITECT Intact PTH Calibrators consists of 6 bottles (4.0 mL each). Calibrator A contains Bis Tris Propane4 de junio de 2014
Preguntas sobre este retiro
¿ARCHITECT Intact PTH Controls consists of 3 x 1 Bottle (8.0 mL each) of PTH (synthetic peptide) in Bis Tris Propane buffer with protein (bovine) stabilizer. Preservatives: sodium azide and ProClin 300. Product Usage: The ARCHITECT Intact PTH assay is an in vitro chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of intact parathyroid hormone (PTH) in human serum and plasma on the ARCHITECT i System. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1672-2014 el 4 de junio de 2014. Peligro reportado: Abbott has confirmed that a performance shift in the ARCHITECT Intact PTH assay has the potential to generate falsely elevated results on patient samples. Results generated with impacted lots may demonstrate a positive shift relative to those generated with previous calibrator and/or control lots. The issue may also impact your established ARCHITECT Intact PTH reference ranges. Abbott ARCHITECT In
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter