Integra Licox Test Kit REF BC10R Rx Only An accessory provided for use with the LCXO2 and AC31 PtO2 Monitors. It provides users with a secondary means to verify the monitor's PtO2 state-of-calibration (but does not provide calibration) when needed by providing a specific electrical signal to the monitor.

Publicado el 11 de junio de 2014 · Inicio del retiro: 16 de abril de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Integra LifeSciences has identified through a small number of complaints that some lot numbers of Licox Test Kits (Catalog Number BC10R) may provide erroneous "calibration test failure" message when they are used with the Licox PtO2 Monitor (LCSO2, or may provide an erroneous out of range result with the Licox CMP Brain Tissue Oxygen and Temperature Monitor (AC31).

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Integra LifeSciences Corp.
Distribuidor
Worldwide Distribution - USA including Illinois, Florida, Maryland, Massachusetts, North Carolina, Pennsylvania, South Dakota and Internationally to Australia, Austria, Belgium, Chile, Finland, Germany, Great Britain, Greece, Israel, Italy and Russia.
Lote
Catalog Number BC10R Lot Nos: 051113, 100913, 140513, 161013, 25013
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States