FDAFDA · Alimentos y medicamentos·N.º oficial Z-1675-2014
Integra Licox Test Kit REF BC10R Rx Only An accessory provided for use with the LCXO2 and AC31 PtO2 Monitors. It provides users with a secondary means to verify the monitor's PtO2 state-of-calibration (but does not provide calibration) when needed by providing a specific electrical signal to the monitor.
Publicado el 11 de junio de 2014 · Inicio del retiro: 16 de abril de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Integra LifeSciences has identified through a small number of complaints that some lot numbers of Licox Test Kits (Catalog Number BC10R) may provide erroneous "calibration test failure" message when they are used with the Licox PtO2 Monitor (LCSO2, or may provide an erroneous out of range result with the Licox CMP Brain Tissue Oxygen and Temperature Monitor (AC31).
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Integra LifeSciences Corp.
- Distribuidor
- Worldwide Distribution - USA including Illinois, Florida, Maryland, Massachusetts, North Carolina, Pennsylvania, South Dakota and Internationally to Australia, Austria, Belgium, Chile, Finland, Germany, Great Britain, Greece, Israel, Italy and Russia.
- Lote
- Catalog Number BC10R Lot Nos: 051113, 100913, 140513, 161013, 25013
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States