***REF C7120***APEX Arthroscopy Tubing Set. Intended for use for joint irrigation and dissention during arthroscopic procedures.
Publicado el 17 de julio de 2013 · Inicio del retiro: 30 de mayo de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and C7122 APEX Arthroscopy Tubing Set, Two Connection, contained a scratch/grooved area on the inside diameter of the pressure fitting that may cause the APEX Irrigation (pump) Console to sense a pressure drop and increase fluid flow to the surgical site.
Qué hacer
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Detalles del retiro
- Fabricante
- Linvatec Corp. dba ConMed Linvatec
- Distribuidor
- Worldwide Distribution-USA (nationwide) including the states of AR, CA, FL, GA, IL, IN, KS, MO, NC, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, and WI, and the countries of Austria; Belgium, Canada, China, Czech Republic, Finland, France, Great Britain, Italy, and Japan.
- Lote
- C7120 Lot's #: 1301184, 1301214, 1301254, 1301284, 1301294, 1302014, 1302044, 1302054, 1302064, 1302074, 13020124, 1302134, 1302144, 1302194, 1302204, 1302214, 1302224, 1302254, 1302264, 1303014, 1303054, 1303064, 1303074, 1303114, 1303124, 1303204, 1303214, 1303264, 1303274, 1303284, 1304014, 1304034, 1304054, 1304154, 1304294, 1304304, and 1305014.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States