FDAFDA · Alimentos y medicamentos·N.º oficial Z-1711-2014

AXIOM Luminos dRFsystem The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.

Publicado el 11 de junio de 2014 · Inicio del retiro: 11 de abril de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

It was determined that under rare environmental conditions (i.e., extremely dry air in combination with a highly isolated floor) electrostatic discharges (ESD) >8 kV may trigger an unintended movement of the AXIOM Luminos dRFsystem which may lead to an emergency situation and pose danger to the patients, operating personnel or to the unit.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Siemens Medical Solutions USA, Inc
Distribuidor
Class II Recall - US Distribution including the states of IL,VA, TN, PA, TX, MN, GA, MI, IN, OH, FL, CA, OR, NY, ME, MO, OK, CO and NJ.
Lote
Model number: 10094200, serial numbers:4332 4291 4323 4430 3061 4331 4238 4297 4201 4024 3098 4252 4372 4354 4077 3050 3108 4327 4342 4251 3053 3114 4156 3145 3096 4043 3181 4208 3090 4143 4051 4199 4401 4405 4151 4373 4276
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States