Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1714-2026

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: CVOR ANGIOGRAPHY PACK-LF, Medline SKU DYNJ42367B

Publicado el 15 de abril de 2026 · Inicio del retiro: 27 de febrero de 2026

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Qué hacer

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Detalles del retiro

Fabricante
Medline Industries, LP
Distribuidor
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
Lote
Medline SKU DYNJ42367B UDI/DI each 10193489211511, UDI/DI case 40193489211512, Lot Numbers: 24FBJ579, 24DBL828, 24DBE055, 24CBG967, 24ABN254, 23LBI540, 23KBF222, 23JBM761, 23JBA270, 23HBY791, 23HBD212, 23GBJ794, 23FBK893, 23EBQ313, 23DBC465.
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.