Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
Publicado el 8 de abril de 2026 · Inicio del retiro: 13 de febrero de 2026
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Medtronic MiniMed, Inc.
- Distribuidor
- Worldwide - US Nationwide distribution including in the states of OR, CA, AZ, NY,TX,FL, OH, MO, TN,NJ, NC, SC, IL, VA, WA, DC, GA, PA, MS,IA, MI, NH, OK, SD, MT, WI, WV, MA, LA, MN, RI, AR, ME, AL, IN, UT, KY, CO, ND, MD, CT, NM, KS, ID, NV, DE, NE, VT, AK, HI, WY, PR, VI and the countries of Taiwan, New Zealand, Australia, Bangladesh, Brunei Darussalam, Cambodia, Indonesia, Malaysia, Maldives, Philippines, Thailand, Vietnam, Singapore, Pakistan, Hong Kong, China, Argentina, Aruba, Barbados, Bermuda, Bolivia, Plurinational State Of, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, Puerto Rico, Trinidad And Tobago, Uruguay, Venezuela, Canada, South Korea, India, Albania, Algeria, Andorra, Armenia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Bulgaria, Canary Islands, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guyana, French Polynes., Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Macedonia, Malawi, Malta, Martinique, Mauritius, Mayotte, Moldavia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Qatar, Reunion, Romania, Russia, San Marino, Saudi Arabia, Serbia, Sierra Leone, Slovakia, Slovenia, South Africa, Spain, St.Pier,Miquel., Sudan, Sweden, Switzerland, Syrian Arab Republic, Tanzania, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Yemen, Zambia, Zimbabwe.
- Lote
- Pump: UDI-DI(GTIN): User Guide Name(Part-Version): Paradigm Real-Time Insulin Pump: 00613994409423, 00613994622600, 613994622747, 613994407733, 00613994407740, 00613994136374, 00613994271600, 00613994106476, 00613994373120, 00613994409478, 00613994312426, 00613994151780, 00885074419831, 00613994622747, 00613994951502, 00613994622648, 00643169026100, 00613994622624, 00613994622556, 00613994402011, 00613994409553, 00613994368485, 00613994368508, 00613994368522, 00613994108739, 00613994368607, 00613994368492, 00613994368515, 00613994123084, 00613994124814, 00613994402523, 00613994402530, 00613994402516, 00613994401502, 00613994402509, 00613994405913, 00613994123206, 00643169688421, 00643169199699, 00643169199675, 00613994409461, 00613994124777, 00643169572492, 00613994267177, 00643169848054, 00613994266897, 00613994310804, 00613994310798, 00613994310774, 00613994310903, 00613994401564, 00613994401663, 00613994402554, 00613994402462, 00613994402592, 00613994402608, 00613994405944, 00613994402455, 00613994402547, 00613994402479, 00763000069292, 00613994402578, 00613994409447, 00643169729438, 00763000069254, 00763000084325, 00613994402486, 00613994405920, 00763000069261, 00613994948595, 00613994929778, 00613994224569, 00613994929808, 00613994930361, 00613994929761, 00643169199705, 00643169199668, 00643169496248, 00643169848016, 00643169848023, 00763000161095,00613994123329, 00613994125026, 00613994247728, 00613994124692, 00613994124739, 00613994266903, 00613994267153, 00613994266880, 00613994267207, 00613994310866, 00613994310811, 00613994310781, 00613994310873, 00613994310613, 00613994310910, 00613994310880, 00613994310767, 00613994375025, 00613994368638, 00613994368553, 00613994368560, 00613994405937, 00613994368539, 00613994368546, 00613994377586, 00613994368591, 00613994368584, 00613994368577, 00613994402585, 00613994402561, 00613994401588, 00613994929815, 00643169221970, 00763000069247, 00763000090654, 00763000640163, 00613994601629, 00643169354975, 00613994401977, 00763000084332, 00643169919389, 00763000161118, 00763000161071: Artwork, CRB D Process, Instr, 522/722, NA(MP6025250-013DOC); Artwork, CRB D Process, Paradigm¿ 522 and 722 Insulin Pumps User(MP6025681-022DOC); Artwork, CRB A Process, Instr, 522/722,Block Mode,TST,Nordic,EN(MP6025681-023DOC); Artwork, CRB D Process, Instr, 522/722 with Enlite, OUS, EN(MP6025646-021DOC); Artwork, CRB D Process, Instr, 522/722, w/Enlite, OUS, EN(MP6025646-022DOC); Artwork, CRB A Process, Instr, 522/722, w/Enlite, OUS, EN(MP6025646-023DOC); Artwork, CRB A Process, Instr, 522/722, w/Enlite, OUS, EN(MP6025646-023DOC); Artwork, New Product, Feature (Form, Fit, or Function), Regulatory, Technical Literature, CRB; IFU, MMT-522/722 PRDGM X22 PUMP W/ENLITE, EN, OUS(MP6025646-024DOC); Artwork, New Product, Feature (Form, Fit, or Function), Regulatory, Technical Literature, CRB; IFU, MMT-522/722 PRDGM X22 PUMP W/ENLITE, EN, OUS(MP6025646-024DOC); Artwork, New Product, Feature (Form, Fit, or Function), Regulatory, Technical Literature, CRB; IFU, MMT-522/722 PRDGM X22 PUMP W/ENLITE, EN, OUS - Nordic cover (MP6025681-024DOC);
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.