GE Healthcare Allia IGS 7 OR, Interventional fluoroscopic x-ray systems

Publicado el 15 de mayo de 2024 · Inicio del retiro: 18 de marzo de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

GE HealthCare has become aware of the potential that adhesive material might not have been applied to some bolts of certain Innova IGS, Discovery IGS, Interact IGS, Allia IGS and Allia Pulse IGS products. This could result in a potential fall of the Detector/Elevator or X-ray tube of the frontal Gantry assembly.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
GE Medical Systems, SCS
Distribuidor
Worldwide distribution - US Nationwide in the states of CA, FL, GA, IN, MA, NC, OK, SC, TN, TX, VA and the countries of Armenia, China, France, Hungary, India, Italy, Japan, Kazakhstan, Lithuania, Mexico, Russia, Spain, Taiwan, Turkey, Vietnam
Lote
UDI/DI 00195278215550, Serial Numbers: D3-23-026, D4-24-002
Clasificación de riesgo
Class II
Área (FDA)
device
País
France