FDAFDA · Alimentos y medicamentos·N.º oficial Z-1809-2014
Model Number(s): PB840 Ventilator (4-840120DIUU-XX). The product is a component of the Puritan Bennett 840 Ventilator: The Backlight Inverter Printed Circuit Board (BLI PCBA) for the 9.4 Display part number 4-079056-00. The PB840 Ventilator System is intended to provide invasive or non-invasive ventilator support and monitoring for infant, pediatric, and adult patients with respiratory failure or respiratory insufficiency. It is used in hospitals and healthcare facilities
Publicado el 25 de junio de 2014 · Inicio del retiro: 4 de junio de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
In the case of a loss of GUI display due to a Backlight Inverter PCBA failure, the ventilator continues to provide uninterrupted ventilatory support at the programmed settings for the patient. However, there is a loss of display and thus there is a necessity to move the patient to another ventilator.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Nellcor Puritan Bennett Inc. (dba Covidien LP)
- Distribuidor
- Worldwide Distribution - USA (nationwide) including the states: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and District of Columbia., and the countries of Canada and Japan.
- Lote
- Display part number 4-079056-00; serial number 4200000001 to 3510012377 that have a 9.4 display. Lots 1032H,1041H,1042H,1043H,1008J,1009J,1010J,1018K,1019K,1020K,1021K,1022K,1023K,1030L,1031L,1032L,1033L,1034L,1035L,
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States