The product is Pathfinder SL, version 1.6.0.3057 software used with the Spacelabs Healthcare Pathfinder SL Holter Analyzer, version 1.6.0. Product Usage: The Holter Analyzer analyzes information collected from ambulatory electrocardiogram recorders worn by patients and used to assist physicians in diagnosis and patient monitoring. In addition, incorrect QTc values are reported after moving the cursor. QTc values are timing events between labeled points of interest in the ECG signal that the physician can mark with the cursor. The software is used to edit a family of events and upon conclusion the software crashes with an error message.
Publicado el 27 de junio de 2012 · Inicio del retiro: 11 de mayo de 2012
Resumen
FDA registra el retiro Z-1831-2012 de dispositivos médicos de Clase II, publicado el 27 de junio de 2012 e iniciado el 11 de mayo de 2012. Empresa responsable: Spacelabs Healthcare, Llc; Estado: Finalizado; Cantidad reportada: 114; Lote: Version 1.6.0. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The firm received reports of the Pathfinder LS software crashing after editing a batch of Holter Analyzer records. The software will stop with an error message requiring a restart. In addition, QTc values will not update after moving the marker.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Spacelabs Healthcare, Llc
- Distribuidor
- Worldwide Distribution - US (nationwide) including the states of Nevada and Vermont, and the countries of AUSTRALIA, AUSTRIA, CANADA, CYPRUS, GERMANY, ICELAND, ISRAEL, POLAND, ROMANIA, SOUTH AFRICA, SWITZERLAND, THAILAND, TURKEY and UNITED KINGDOM.
- Lote
- Version 1.6.0.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDASpacelabs Healthcare G2 Clinical Access Mislabeled CDs for PN: 063-1829-11 Rev A and Update Kit, PN:2 de abril de 2014
- FDAARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C. Model 96281 is intended for use with either5 de febrero de 2014
- FDASpacelabs Healthcare qube Compact Patient Monitor, Model 91390. It is a patient monitor with arrhythmia27 de noviembre de 2013
- FDASpacelabs Healthcare elance Vital Signs Monitor, Model 93300, with Option S. It is a patient monitor with20 de noviembre de 2013
- FDAPhilips Multi Diagnost Eleva II with swivel cable Product Codes; 708032, 708036 Multi Diagnost Eleva with27 de junio de 2012
- FDASARNS HIGH-FLOW AORTIC ARCH CANNULA: 5.2 MM (16 FR) 00 WITH 3/8" CONNECTOR, 7.5" (19 CM) LONG. The Sarns High27 de junio de 2012
Preguntas sobre este retiro
¿The product is Pathfinder SL, version 1.6.0.3057 software used with the Spacelabs Healthcare Pathfinder SL Holter Analyzer, version 1.6.0. Product Usage: The Holter Analyzer analyzes information collected from ambulatory electrocardiogram recorders worn by patients and used to assist physicians in diagnosis and patient monitoring. In addition, incorrect QTc values are reported after moving the cursor. QTc values are timing events between labeled points of interest in the ECG signal that the physician can mark with the cursor. The software is used to edit a family of events and upon conclusion the software crashes with an error message. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1831-2012 el 27 de junio de 2012. Peligro reportado: The firm received reports of the Pathfinder LS software crashing after editing a batch of Holter Analyzer records. The software will stop with an error message requiring a restart. In addition, QTc values will not update after moving the marker.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter