FDAFDA · Alimentos y medicamentos·N.º oficial Z-1831-2012
The product is Pathfinder SL, version 1.6.0.3057 software used with the Spacelabs Healthcare Pathfinder SL Holter Analyzer, version 1.6.0. Product Usage: The Holter Analyzer analyzes information collected from ambulatory electrocardiogram recorders worn by patients and used to assist physicians in diagnosis and patient monitoring. In addition, incorrect QTc values are reported after moving the cursor. QTc values are timing events between labeled points of interest in the ECG signal that the physician can mark with the cursor. The software is used to edit a family of events and upon conclusion the software crashes with an error message.
Publicado el 27 de junio de 2012 · Inicio del retiro: 11 de mayo de 2012
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The firm received reports of the Pathfinder LS software crashing after editing a batch of Holter Analyzer records. The software will stop with an error message requiring a restart. In addition, QTc values will not update after moving the marker.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Spacelabs Healthcare, Llc
- Distribuidor
- Worldwide Distribution - US (nationwide) including the states of Nevada and Vermont, and the countries of AUSTRALIA, AUSTRIA, CANADA, CYPRUS, GERMANY, ICELAND, ISRAEL, POLAND, ROMANIA, SOUTH AFRICA, SWITZERLAND, THAILAND, TURKEY and UNITED KINGDOM.
- Lote
- Version 1.6.0.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States