Nextgen Complete Knee Solution Replacement Jaw for Use with Femoral and Provisional Impactor/Extractor, REF 00-5901-026-40, Non-Sterile Zimmer, Warsaw Indiana The instrument is attached to the femoral implant or provisional to aid in positioning and inserting the components.
Publicado el 8 de agosto de 2012 · Inicio del retiro: 30 de abril de 2012
Resumen
FDA registra el retiro Z-2092-2012 de dispositivos médicos de Clase II, publicado el 8 de agosto de 2012 e iniciado el 30 de abril de 2012. Empresa responsable: Zimmer Inc.; Estado: Finalizado; Cantidad reportada: 65 units; Lote: Part 00-5901-026-40 Lot 62011774, 61870067, and 61813699 Initially the 00-5901-026-40 replacement jaws will remain available for order until all of the alternative instruments are in place, however they will also be required to be returned. The final lot numbers and manufacture dates will be provided in a subsequent status update. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The design of the spring clip for the jaw assembly was updated in March 2011 due to fractures of the spring clip. Zimmer has received complaints for the spring clip fracturing on the updated design. The spring clip may fracture or fall off of the instrument during use, and a device fragment may be left in the surgical site. The risk of disassembly or fracture occurs when the instrument is typicall
Qué hacer
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Zimmer Inc.
- Distribuidor
- Worldwide Distribution - USA (nationwide) and the countries of Singapore, China, Hong Kong, Canada, Australia, Japan, Europe, Germany, India, and Finland.
- Lote
- Part 00-5901-026-40 Lot 62011774, 61870067, and 61813699 Initially the 00-5901-026-40 replacement jaws will remain available for order until all of the alternative instruments are in place, however they will also be required to be returned. The final lot numbers and manufacture dates will be provided in a subsequent status update.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDANextgen Complete Knee Solution Femoral and Provisional Impactor/ Extractor, REF 00-5901-026-00, Non-Sterile8 de agosto de 2012
- FDAArm and Hammer Spinbrush Pro Clean Sonic Powered Toothbrush Made in China. Manufactured for Church and Dwight8 de agosto de 2012
- FDAArm and Hammer Spinbrush Dual Action Powered Toothbrush Spinbrush Pro Whitening Made in China. Manufactured8 de agosto de 2012
- FDAArm and Hammer Spinbrush Pro Clean Sonic Color Wear Powered Toothbrush Made in China. Manufactured for Church8 de agosto de 2012
- FDAArm and Hammer Spinbrush Dual Action Powered Toothbrush Pro Clean Re-Charge Made in China. Manufactured for8 de agosto de 2012
- FDAArm and Hammer Spinbrush Pro White Sonic Crest Powered Toothbrush Made in China. Manufactured for Church and8 de agosto de 2012
Preguntas sobre este retiro
¿Nextgen Complete Knee Solution Replacement Jaw for Use with Femoral and Provisional Impactor/Extractor, REF 00-5901-026-40, Non-Sterile Zimmer, Warsaw Indiana The instrument is attached to the femoral implant or provisional to aid in positioning and inserting the components. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2092-2012 el 8 de agosto de 2012. Peligro reportado: The design of the spring clip for the jaw assembly was updated in March 2011 due to fractures of the spring clip. Zimmer has received complaints for the spring clip fracturing on the updated design. The spring clip may fracture or fall off of the instrument during use, and a device fragment may be left in the surgical site. The risk of disassembly or fracture occurs when the instrument is typicall
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit