The Reliance 130 and 130L Cart Washer/Disinfector is a high capacity mechanical washer intended for use in the washing, low-level disinfecting, and drying of carts, utensils, beds and other miscellaneous reusable items used in the care of patients at healthcare facilities / Crated for shipping.
Publicado el 8 de agosto de 2012 · Inicio del retiro: 14 de julio de 2012
Resumen
FDA registra el retiro Z-2122-2012 de dispositivos médicos de Clase II, publicado el 8 de agosto de 2012 e iniciado el 14 de julio de 2012. Empresa responsable: Steris Corporation; Estado: Finalizado; Cantidad reportada: 367; Lote: The codes for the product are Julian date-coded. They follow the following pattern: AABBBCCDDD AA is a two digit code representing the manufacturing facility number BBB is a three digit code representing the day of the year that the device was manufactured CC is a two digit code representing the Julian year that the device was manufactured DDD is a three digit code representing the number of the number of the device manufactured that day. For example, a serial number of 3606499005 indicates that the device was manufactured in facility number 36, on the 64th day of the year of 1999, and the device was the 5th device manufactured that day. The serial number ranges for the devices being recalled are: RELIANCE® 130: 3606499005 3626709006 RELIANCE® 130L: 3629299006 3628809011. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
To ensure that users are properly operating the Reliance 130 and 130L Cart Washers, a door close confirmation button and door open delay will be installed on your cart washer.
Qué hacer
Voluntary: Firm initiated
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Steris Corporation
- Distribuidor
- Worldwide Distribution -- USA (nationwide) including the states of : AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY. and the countries of : Australia, South Korea, France, Italy, Lebanon, Russia, Saudi Arabia, Spain, UAE, United Kingdom and Canada.
- Lote
- The codes for the product are Julian date-coded. They follow the following pattern: AABBBCCDDD AA is a two digit code representing the manufacturing facility number BBB is a three digit code representing the day of the year that the device was manufactured CC is a two digit code representing the Julian year that the device was manufactured DDD is a three digit code representing the number of the number of the device manufactured that day. For example, a serial number of 3606499005 indicates that the device was manufactured in facility number 36, on the 64th day of the year of 1999, and the device was the 5th device manufactured that day. The serial number ranges for the devices being recalled are: RELIANCE® 130: 3606499005 3626709006 RELIANCE® 130L: 3629299006 3628809011
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML02011721 de octubre de 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML02011521 de octubre de 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML02011621 de octubre de 2026
- FDAVERIFY Biological Steam Test Pack Model/Catalog Number: S306914 de octubre de 2026
- FDAPerkin Elmer Genetic Screening Processor (GSP) Product Code: 2021-0010 Product Usage: The GSP instrument is a8 de agosto de 2012
- FDAIntegra® Advansys Medial Lisfranc Plate (MLP), Integra, 311 Enterprise Drive, Plainsboro, NJ 08536 The8 de agosto de 2012
Preguntas sobre este retiro
¿The Reliance 130 and 130L Cart Washer/Disinfector is a high capacity mechanical washer intended for use in the washing, low-level disinfecting, and drying of carts, utensils, beds and other miscellaneous reusable items used in the care of patients at healthcare facilities / Crated for shipping. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2122-2012 el 8 de agosto de 2012. Peligro reportado: To ensure that users are properly operating the Reliance 130 and 130L Cart Washers, a door close confirmation button and door open delay will be installed on your cart washer.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated