Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2198-2013

ProTime Microcoagulation System; PROTIME CUVETTE AND PROTIME3 CUVETTE; Store the foil-pouched cuvettes refrigerated 2-8C (36 - 36F); MATERIALS PROVIDED; ProTime cuvettes with Tenderlett Plus; or; Protime3 cuvettes with Tenderlett Plus LV; Product Instructions; Antiseptic, gauze (with self-test package only); International Technidyne Corporation; 8 Olsen Avenue; Edison, NJ 08820 USA; a subsidiary of Thoratec Corporation The ProTime Microcoagulation System consists of a portable, battery operated instrument and disposable cuvette for quantitative determination of prothrombin time (PT) from fingerstick whole blood or anticoagulant-free venous whole blood. There are two different types of cuvettes available for use with the ProTime Microcoagulation System: a ProTime cuvette and a ProTime3 cuvette. The ProTime cuvette is black color coded and utilizes three channels for the PT assays and two channels for integral controls. It requires approximately 65 ul of blood (approximately 3 drops). The ProTime3 cuvette is blue color coded and utilizes one channel for PT assay and two channels for integral controls. The ProTime3 cuvette requires approximately 27 ul of blood (approximately 1 large drop). The ProTime Microcoagulation System is intended for professional use in the management of patients treated with oral anticoagulants or for patient self-testing.

Publicado el 25 de septiembre de 2013 · Inicio del retiro: 16 de julio de 2013

Resumen

FDA registra el retiro Z-2198-2013 de dispositivos médicos de Clase II, publicado el 25 de septiembre de 2013 e iniciado el 16 de julio de 2013. Empresa responsable: Assuramed; Estado: Finalizado; Cantidad reportada: 1) 72 units; 2) 56 units; Lote: *** 1) Product Number: PRO3-25; Lots: All Lots; Independence Medical Catalog Number: 55L2101002; *** 2) Product Number: PRO5-25; Lots: All Lots; Independence Medical Catalog Number: ICPRO525. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The products may have been stored at temperatures outside their required storage conditions.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Assuramed
Distribuidor
Nationwide Distribution including AR, AZ, CA, CO, FL, IL, KY, LA, MI, NJ, NY, OH,OK, OR, SC, TX, UT, and WA.
Lote
*** 1) Product Number: PRO3-25; Lots: All Lots; Independence Medical Catalog Number: 55L2101002; *** 2) Product Number: PRO5-25; Lots: All Lots; Independence Medical Catalog Number: ICPRO525
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿ProTime Microcoagulation System; PROTIME CUVETTE AND PROTIME3 CUVETTE; Store the foil-pouched cuvettes refrigerated 2-8C (36 - 36F); MATERIALS PROVIDED; ProTime cuvettes with Tenderlett Plus; or; Protime3 cuvettes with Tenderlett Plus LV; Product Instructions; Antiseptic, gauze (with self-test package only); International Technidyne Corporation; 8 Olsen Avenue; Edison, NJ 08820 USA; a subsidiary of Thoratec Corporation The ProTime Microcoagulation System consists of a portable, battery operated instrument and disposable cuvette for quantitative determination of prothrombin time (PT) from fingerstick whole blood or anticoagulant-free venous whole blood. There are two different types of cuvettes available for use with the ProTime Microcoagulation System: a ProTime cuvette and a ProTime3 cuvette. The ProTime cuvette is black color coded and utilizes three channels for the PT assays and two channels for integral controls. It requires approximately 65 ul of blood (approximately 3 drops). The ProTime3 cuvette is blue color coded and utilizes one channel for PT assay and two channels for integral controls. The ProTime3 cuvette requires approximately 27 ul of blood (approximately 1 large drop). The ProTime Microcoagulation System is intended for professional use in the management of patients treated with oral anticoagulants or for patient self-testing. fue retirado del mercado?

Sí. FDA publicó el retiro Z-2198-2013 el 25 de septiembre de 2013. Peligro reportado: The products may have been stored at temperatures outside their required storage conditions.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter