FDAFDA · Alimentos y medicamentos·N.º oficial Z-2213-2024

The D*Clot® HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system consists of an outer sheath through which a rotatable radiopaque spiral shaft extends ending in an eccentric blunt distal tip. Once activated, the shaft rotates at approximately 10.000 RPMs to break thrombus into smaller particles, which can be simultaneously aspirated using a dedicated vacuum syringe. The catheter is a stand-alone system and is powered by a 4.5 VDC alkaline battery pack in the control unit.

Publicado el 3 de julio de 2024 · Inicio del retiro: 1 de mayo de 2024

Resumen

FDA registra el retiro Z-2213-2024 de dispositivos médicos de Clase II, publicado el 3 de julio de 2024 e iniciado el 1 de mayo de 2024. Empresa responsable: Mermaid Medical A/S; Estado: En curso; Cantidad reportada: 190 devices; Lote: Unique Device Identifier: 15711055004333; Lot codes: S2306001; Model Number: T10106065. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Internal product testing failed bubble leak testing (ASTM F2096), which indicates the sterility barrier may not be fully intact.

Qué hacer

Voluntary: Firm initiated · E-Mail

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Mermaid Medical A/S
Distribuidor
Worldwide - US Nationwide distribution in the states of OH, MO, CO, NC, TX, MA, GA, FL, IL, IN, MD, PA, VA, and the country of Chile.
Lote
Unique Device Identifier: 15711055004333; Lot codes: S2306001; Model Number: T10106065
Clasificación de riesgo
Class II
Área (FDA)
device
País
Denmark

Retiros relacionados

Preguntas sobre este retiro

¿The D*Clot® HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system consists of an outer sheath through which a rotatable radiopaque spiral shaft extends ending in an eccentric blunt distal tip. Once activated, the shaft rotates at approximately 10.000 RPMs to break thrombus into smaller particles, which can be simultaneously aspirated using a dedicated vacuum syringe. The catheter is a stand-alone system and is powered by a 4.5 VDC alkaline battery pack in the control unit. fue retirado del mercado?