The D*Clot® HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system consists of an outer sheath through which a rotatable radiopaque spiral shaft extends ending in an eccentric blunt distal tip. Once activated, the shaft rotates at approximately 10.000 RPMs to break thrombus into smaller particles, which can be simultaneously aspirated using a dedicated vacuum syringe. The catheter is a stand-alone system and is powered by a 4.5 VDC alkaline battery pack in the control unit.
Publicado el 3 de julio de 2024 · Inicio del retiro: 1 de mayo de 2024
Resumen
FDA registra el retiro Z-2213-2024 de dispositivos médicos de Clase II, publicado el 3 de julio de 2024 e iniciado el 1 de mayo de 2024. Empresa responsable: Mermaid Medical A/S; Estado: En curso; Cantidad reportada: 190 devices; Lote: Unique Device Identifier: 15711055004333; Lot codes: S2306001; Model Number: T10106065. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Internal product testing failed bubble leak testing (ASTM F2096), which indicates the sterility barrier may not be fully intact.
Qué hacer
Voluntary: Firm initiated · E-Mail
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Mermaid Medical A/S
- Distribuidor
- Worldwide - US Nationwide distribution in the states of OH, MO, CO, NC, TX, MA, GA, FL, IL, IN, MD, PA, VA, and the country of Chile.
- Lote
- Unique Device Identifier: 15711055004333; Lot codes: S2306001; Model Number: T10106065
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Denmark
Retiros relacionados
- FDAProduct not distributed in the US. Implantable Cardioverter Defibrillators: a. VIVA QUAD XT CRT-D, Model3 de julio de 2024
- FDASoftware version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPoint3 de julio de 2024
- FDACROME VR MRI SureScan, Model Number DVPC3D1, Implantable Cardioverter Defibrillator3 de julio de 2024
- FDA2nd Generation CentriMag Primary Console, REF: 201-90401, 201-90701, 201-90411, 201-90421, 201-30300,3 de julio de 2024
- FDACOBALT XT VR MRI SureScan, Model Number DVPA2D1, Implantable Cardioverter Defibrillator3 de julio de 2024
- FDANovaGuide 2 Intelligent Ultrasound, REFs: NSC-NVGSYS2 & NSC-NVGSYS2-CA3 de julio de 2024
Preguntas sobre este retiro
¿The D*Clot® HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system consists of an outer sheath through which a rotatable radiopaque spiral shaft extends ending in an eccentric blunt distal tip. Once activated, the shaft rotates at approximately 10.000 RPMs to break thrombus into smaller particles, which can be simultaneously aspirated using a dedicated vacuum syringe. The catheter is a stand-alone system and is powered by a 4.5 VDC alkaline battery pack in the control unit. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2213-2024 el 3 de julio de 2024. Peligro reportado: Internal product testing failed bubble leak testing (ASTM F2096), which indicates the sterility barrier may not be fully intact.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · E-Mail