Sterile convenience kits: a. Medline Retrograde PK-22025203-LF, Pack #DYNJ47873K, containing a Pure Pouch containing component Sterile Curved Kelly Hemostat, Reorder #66175. b. Medline Cath Lab Pack, Pack #DYNJ56141A, containing a Pure Pouch containing component Medline DYND0400 O.R. Scissors, S/B, 5.5 Sterile. c. Medline Sterile Loop Insertion Tray, Pack DYNJ62843A, containing a Pure Pouch containing Medline Reorder #DYNJ04048, Kelly Forceps/Std Str Sterile. d. Medline Radiology Minor MVH, Pack #DYNJ63555C, containing a Pure Pouch containing component Sterile Curved Kelly Hemostat, Reorder #66175. e. Medline Basic Pack, Pack #DYNJ80649B, containing a Pure Pouch containing component Medline DYND0400 O.R. Scissors, S/B, 5.5 Sterile.

Publicado el 10 de julio de 2024 · Inicio del retiro: 14 de marzo de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Surgical convenience kits containing Pure Pouch component that has weak seals which may result in a breach in sterility of the contents if the seal fails.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
MEDLINE INDUSTRIES, LP - Northfield
Distribuidor
US Nationwide distribution in the states of AL, CA, CO, CT, FL, IL, IN, MD, MN, MT, NC, NJ, NV, NY, OH, PA, SC, TN, TX, VA, WA, WI, and WY.
Lote
a. Pack #DYNJ47873K - lot numbers 23HBK868, exp. 8/31/2027, 23HBC082exp. 2/29/2028, 23GBC462, exp. 1/31/2028, and 23DBK792, exp. 11/30/2027, UDI-DI each 10193489466812 and UDI-DI case-40193489466813. b. Pack #DYNJ56141A - lot numbers 23IBK595, exp. 1/31/2025, UDI-DI each-10193489555585 and UDI-DI case-40193489555586. c. Pack DYNJ62843A - lot numbers 24ABP043, exp. 12/31/2024, and 23IBM893, exp. 3/31/2025, UDI-DI each-10195327001766 and UDI-DI case-40195327001224. d. Pack #DYNJ63555C - lot numbers 24ALA613, exp. 3/31/2025; 23KLA748, 1/31/2025; 23KLA017, exp. 1/31/2025; 23JLA853, exp. 1/31/2025; 23JLA232. exp. 1/31/2025; 23ILB006, exp. 1/31/2025; 23ILA201, exp. 12/31/2024; and 23HLA429, exp. 12/31/2024; UDI-DI each-10193489816525 and UDI-DI case-40193489816526. e. Pack #DYNJ80649B - lot number 23JBG046, exp. 5/31/2026, UDI-DI each-10195327269944 and UDI-DI case-40195327269945.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States