FDAFDA · Alimentos y medicamentos·N.º oficial Z-2255-2024

BD Pyxis, Name/REF: MEDBANK MINI 1FH-1FM/ 169-137, MEDBANK MINI 1FH-2HH MEDPASS/ 169-151, MEDBANK MINI 1HH-1FH-1HM/ 169-101, MEDBANK MINI 1HH-1HM-1FM/ 169-102, MEDBANK MINI 1HH-3HM/ 169-103, MEDBANK MINI 2FH/ 169-104, MEDBANK MINI 2FM/ 169-105, MEDBANK MINI 2HH-1FH/ 169-106, MEDBANK MINI 2HH-1FM/ 169-107, MEDBANK MINI 2HH-1FM MEDPASS/ 169-152, MEDBANK MINI 2HH-2HM/ 169-108, MEDBANK MINI 2HM-1FM/ 169-109, MEDBANK MINI 3HH-1FG/ 169-110, MEDBANK MINI 3HH-1HM/ 169-111, MEDBANK MINI 4HH/ 169-112, MEDBANK MINI 4HH MEDPASS/ 169-150, MEDBANK MINI 4HM/ 169-113, MEDBANK MINI CR-2HH-1FH-P/ 169-114, MEDBANK MINI CR-4HH-P/ 169-115, MB MINI MAIN CR/ 169-163, MB MINI MAIN W/CUBIE DRWS/ 169-159, MB MINI MAIN W/FH FALSE DRW/ 169-162, MB MINI MAIN W/HH FALSE DRW/ 169-161, MB MINI MAIN W/MATRIX DRWS / 169-160, MB MINI MAIN W/MP/ 169-164 Used with the following user guides: BD Pyxis CUBIE Replenishment Station User Guide (DIR 10000414989), BD Pyxis MedBank Cabinet User Guide (DIR 10000414987), BD Pyxis MedBank Installation and Upgrade Guide (DIR 10000414990)

Publicado el 10 de julio de 2024 · Inicio del retiro: 13 de mayo de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Multiple socket-outlet power strips that shipped with automated dispensing cabinet devices have not been certified to International Electrotechnical Commission (IEC) electrical standards when used as a system with automated dispensing cabinet devices, so use may expose user to leakage currents that could cause varying degrees of harm from a minor shock to permanent impairment.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
CareFusion 303, Inc.
Distribuidor
US Nationwide distribution including in the states of NY, CA, MO, AZ, CO, KS, IL, FL, OK, OH, TN, GA, NM, NV, NE, TX, NC, KY, WI, SC, IA, MT, SD, VA, NJ, PA, AK, MS, MI, WA, UT, ND, MA, MN, OR, IN, WV, AL, ID, LA, CT, WY, MD, NH, AR, HI, DC.
Lote
UDI-DI: 10885403512544. Devices Manufactured November 15, 2012 - July, 25 2023 User Guide, DIR 10000414989/DIR 10000414987/DIR 10000414990, Version 00.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States