ANSPACH - Fluted Barrel; 8MM x 12.5MM REF L-8BA-G1 Cutting shaping bone including spine and cranium.

Publicado el 5 de agosto de 2015 · Inicio del retiro: 13 de abril de 2015

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

G1 Dissection Tools are not compatible with all attachments listed on the labeling; G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
The Anspach Effort, Inc.
Distribuidor
Worldwide Distribution - US including AZ, CA, CO, FL, GA, IL, KY, LA, MN, MO, NC, NJ, NY, OH, OR, PA, TX, and Internationally to Canada and China.
Lote
REF L-8BA-G1 ; ALL lot numbers
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States