FDAFDA · Alimentos y medicamentos·N.º oficial Z-2288-2026
Philips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Publicado el 10 de junio de 2026 · Inicio del retiro: 7 de diciembre de 2023
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Qué hacer
Letter
Detalles del retiro
- Fabricante
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Distribuidor
- US Nationwide distribution.
- Lote
- UDI (01)00884838099265(21)104, (01)00884838085350(21)221, (01)00884838099265(21)153, (01)00884838099265(21)46, (01)00884838099265(21)28, (01)00884838085350(21)250, (01)00884838099265(21)72, (01)00884838099265(21)208, (01)00884838099265(21)41, (01)00884838099265(21)122, (01)00884838099265(21)147, (01)00884838085350(21)225, (01)00884838099265(21)262, (01)00884838099265(21)112, (01)00884838099265(21)110, (01)00884838099265(21)207, (01)00884838099265(21)68, (01)00884838099265(21)177, (01)00884838099265(21)29, (01)00884838099265(21)71, (01)00884838099265(21)37, (01)00884838099265(21)175, (01)00884838099265(21)148, (01)00884838099265(21)205, , (01)00884838099265(21)229, (01)00884838099265(21)228, (01)00884838085350(21)277, (01)00884838085350(21)115, (01)00884838099265(21)157, (01)00884838085350(21)163, (01)00884838099265(21)268, (01)00884838085350(21)75, (01)00884838085350(21)132, (01)00884838085350(21)147, (01)00884838085350(21)208, (01)00884838085350(21)104, (01)00884838085350(21)244, (01)00884838085350(21)234, (01)00884838099265(21)53, (01)00884838099265(21)259, (01)00884838099265(21)55, (01)00884838099265(21)235, (01)00884838099265(21)298, (01)00884838085350(21)70, (01)00884838099265(21)233, (01)00884838099265(21)141, (01)00884838099265(21)289, (01)00884838099265(21)249, (01)00884838099265(21)84, (01)00884838099265(21)111, (01)00884838099265(21)227.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Netherlands