FDAFDA · Alimentos y medicamentos·N.º oficial Z-2323-2024
Welch Allyn CP150 Electrocardiograph with Spirometry Option, Product Code: CP150A-1ENB. The Electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function.
Publicado el 17 de julio de 2024 · Inicio del retiro: 3 de junio de 2024
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
During manufacturing, the step to install an Electromagnetic Interference (EMI) Absorber was omitted. The purpose of the EMI absorber is to prevent EMI leakage that might potentially affect other devices in the vicinity or the device itself may be vulnerable to EMI resulting in device malfunction or unavailability.
Qué hacer
Voluntary: Firm initiated · Telephone
Detalles del retiro
- Fabricante
- Baxter Healthcare Corporation
- Distribuidor
- Distribution to Illinois.
- Lote
- UDI/DI: 00732094177640. Product Code: CP150A-1ENB. Serial Number: 100013892024
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.