da Vinci X Surgical System, Model IS4200. Product Usage: The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Mode IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
Publicado el 11 de julio de 2018 · Inicio del retiro: 7 de marzo de 2018
Resumen
FDA registra el retiro Z-2341-2018 de dispositivos médicos de Clase II, publicado el 11 de julio de 2018 e iniciado el 7 de marzo de 2018. Empresa responsable: Intuitive Surgical, Inc.; Estado: Finalizado; Cantidad reportada: 108 ACJ's on 35 surgical systems; Lote: System S/N SL0008, ACJ S/N's P117154256, P117154263, P117154281, P117154286; System S/N SL0012, ACJ S/N's P117154260, P117154297, P117154298, P117154308; System S/N SL0013, ACJ S/N's P117154251, P117154255, P117154257, P117154305; System S/N SL0015, ACJ S/N's P117154274, P117154276, P117154289, P117154290; System S/N SL0016, ACJ S/N's P117154259, P117154291, P117154306, P117154307; System S/N SL0017, ACJ S/N's P117154261, P117154269, P117154277, P117154296; System S/N SL0018, ACJ S/N's P117194009, P117194017, P117194037; System S/N SL0019, ACJ S/N's P117154267, P117154283, P117154294, P117154302; System S/N SL0020, ACJ S/N's P117194008, P117194050, P117194060, P117194078; System S/N SL0021, ACJ S/N's P117194063, P117194064, P117194081, P117194083; System S/N SL0028, ACJ S/N's P117154279, P117154285, P117154292, P117154300; System S/N SL0034, ACJ S/N's P117154264, P117154275, P117154304; System S/N SL0035, ACJ S/N's P117194016, P117194019, P117194070, P117194093; System S/N SL0036, ACJ S/N's P117194025, P117194073, P117241990; System S/N SL0037, ACJ S/N's P117194006, P117194036, P117194056, P117194058 System S/N SL0038, ACJ S/N's P117241971, P117241993; System S/N SL0039, ACJ S/N's P117194013, P117194024, P117194031, P117194032; System S/N SL0040, ACJ S/N's P117194046, P117194061, P117194088, P117194094; System S/N SL0041, ACJ S/N's P117194027, P117194040, P117194043, P117194049; System S/N SL0042, ACJ S/N's P117154288, P117194015, P117194052, P117194090; System S/N SL0043, ACJ S/N P117241988; System S/N SL0044, ACJ S/N's P117194007, P117194085, P117241978; System S/N SL0045, ACJ S/N's P117241965, P117241970, P117241982, P117241984; System S/N SL0046, ACJ S/N's P117194014, P117241964, P117241976, P117241980; System S/N SL0047, ACJ S/N's P117194020, P117194048, P117194067, P117194074 System S/N SL0048, ACJ S/N's P117194004, P117194072, P117200171; System S/N SL0052, ACJ S/N's P117241977, P117241979, P117241986, P117241991; System S/N SL0053, ACJ S/N P117241966; System S/N SL0055, ACJ S/N P117283217; System S/N SL0056, ACJ S/N P117283216; System S/N SL0060, ACJ S/N P117241972; System S/N SL0063, ACJ S/N P117241968; System S/N SL0070, ACJ S/N's P117194028, P117194092; System S/N SL0071, ACJ S/N's P117194018, P117283218; and System S/N SL0073. ACJ S/N P117283214. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Specific da Vinci X Surgical Systems (IS4200) were shipped with defective Arm Controller Joint (ACJ) boards that were found to have a high rate of failure due to a defective component.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Intuitive Surgical, Inc.
- Distribuidor
- US Nationwide AR, CA, FL, GA, IL, IA, MN, MS, NE, OK, TN, and TX. Foreign distribution was made to Australia, Austria, Finland, France, Germany, Italy, Netherlands, Spain, and United Kingdom.
- Lote
- System S/N SL0008, ACJ S/N's P117154256, P117154263, P117154281, P117154286; System S/N SL0012, ACJ S/N's P117154260, P117154297, P117154298, P117154308; System S/N SL0013, ACJ S/N's P117154251, P117154255, P117154257, P117154305; System S/N SL0015, ACJ S/N's P117154274, P117154276, P117154289, P117154290; System S/N SL0016, ACJ S/N's P117154259, P117154291, P117154306, P117154307; System S/N SL0017, ACJ S/N's P117154261, P117154269, P117154277, P117154296; System S/N SL0018, ACJ S/N's P117194009, P117194017, P117194037; System S/N SL0019, ACJ S/N's P117154267, P117154283, P117154294, P117154302; System S/N SL0020, ACJ S/N's P117194008, P117194050, P117194060, P117194078; System S/N SL0021, ACJ S/N's P117194063, P117194064, P117194081, P117194083; System S/N SL0028, ACJ S/N's P117154279, P117154285, P117154292, P117154300; System S/N SL0034, ACJ S/N's P117154264, P117154275, P117154304; System S/N SL0035, ACJ S/N's P117194016, P117194019, P117194070, P117194093; System S/N SL0036, ACJ S/N's P117194025, P117194073, P117241990; System S/N SL0037, ACJ S/N's P117194006, P117194036, P117194056, P117194058 System S/N SL0038, ACJ S/N's P117241971, P117241993; System S/N SL0039, ACJ S/N's P117194013, P117194024, P117194031, P117194032; System S/N SL0040, ACJ S/N's P117194046, P117194061, P117194088, P117194094; System S/N SL0041, ACJ S/N's P117194027, P117194040, P117194043, P117194049; System S/N SL0042, ACJ S/N's P117154288, P117194015, P117194052, P117194090; System S/N SL0043, ACJ S/N P117241988; System S/N SL0044, ACJ S/N's P117194007, P117194085, P117241978; System S/N SL0045, ACJ S/N's P117241965, P117241970, P117241982, P117241984; System S/N SL0046, ACJ S/N's P117194014, P117241964, P117241976, P117241980; System S/N SL0047, ACJ S/N's P117194020, P117194048, P117194067, P117194074 System S/N SL0048, ACJ S/N's P117194004, P117194072, P117200171; System S/N SL0052, ACJ S/N's P117241977, P117241979, P117241986, P117241991; System S/N SL0053, ACJ S/N P117241966; System S/N SL0055, ACJ S/N P117283217; System S/N SL0056, ACJ S/N P117283216; System S/N SL0060, ACJ S/N P117241972; System S/N SL0063, ACJ S/N P117241968; System S/N SL0070, ACJ S/N's P117194028, P117194092; System S/N SL0071, ACJ S/N's P117194018, P117283218; and System S/N SL0073. ACJ S/N P117283214.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAda Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-1522 de julio de 2026
- FDADa Vinci ASSY, PSS, SP1098, Part Number: 3806011 de julio de 2026
- FDABrand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/A10 de junio de 2026
- FDAIntuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-0613 de mayo de 2026
- FDAGE Healthcare CARESCAPE Monitor B65011 de julio de 2018
- FDASafety Sideport Knife 1.0mm 45®, REF 378210 Product Usage: Indicated to make stab incisions during ophthalmic11 de julio de 2018
Preguntas sobre este retiro
¿da Vinci X Surgical System, Model IS4200. Product Usage: The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Mode IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2341-2018 el 11 de julio de 2018. Peligro reportado: Specific da Vinci X Surgical Systems (IS4200) were shipped with defective Arm Controller Joint (ACJ) boards that were found to have a high rate of failure due to a defective component.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter