Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.
Publicado el 31 de julio de 2024 · Inicio del retiro: 4 de junio de 2024
Resumen
FDA registra el retiro Z-2414-2024 de dispositivos médicos de Clase II, publicado el 31 de julio de 2024 e iniciado el 4 de junio de 2024. Empresa responsable: Beckman Coulter, Inc.; Estado: En curso; Cantidad reportada: 143 devices; Lote: Serial numbers: 300117, 300116, 300137, 300138, 300139, 300129, 300141, 300143, 300142, 300151, 300152, 300128, 300156, 300153, 300154, 300135, 300164, 300180, 300155, 300167, 300166, 300158, 300171, 300179, 300169, 300182, 300176, 300183, 300198, 300192, 300133, 300188, 300161, 300134, 300190, 300209, 300212, 300211, 300149, 300148, 300150, 300130, 300172, 300197, 300160, 300189, 300170, 300181, 300195, 300194, 300224, 300222, 300168, 300210, 300233, 300207, 300205, 300221, 300203, 300240, 300199, 300200, 300136, 300196, 300178, 300220, 300208, 300126, 300213, 300206, 300219, 300214, 300217, 300215, 300229, 300255, 300257, 300202, 300254, 300223, 300258, 300259, 300218, 300201, 300234, 300237, 300249, 300260, 300238, 300243, 300261, 300231, 300191, 300131, 300236, 300264, 300263, 300193, 300239, 300230, 300235, 300245, 300216, 300247, 300242, 300226, 300270, 300271, 300269, 300174, 300251, 300241, 300232, 300204, 300227, 300228, 300256, 300266, 300281, 300282, 300253, 300244, 300272, 300246, 300145, 300123, 300147, 300297, 300157, 300275, 300284, 300286, 300248, 300273, 300252, 300144, 300146, 300165, 300177, 300173, 300250, 300268, and 300140; UDI-DI 15099590732103. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed patient results.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Beckman Coulter, Inc.
- Distribuidor
- Worldwide - US Nationwide distribution in the states of FL, IN, KS, NH, PA, and TX. The countries of Australia, Austria, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, Province of China, and United Kingdom of Great Britain and Northern Ireland.
- Lote
- Serial numbers: 300117, 300116, 300137, 300138, 300139, 300129, 300141, 300143, 300142, 300151, 300152, 300128, 300156, 300153, 300154, 300135, 300164, 300180, 300155, 300167, 300166, 300158, 300171, 300179, 300169, 300182, 300176, 300183, 300198, 300192, 300133, 300188, 300161, 300134, 300190, 300209, 300212, 300211, 300149, 300148, 300150, 300130, 300172, 300197, 300160, 300189, 300170, 300181, 300195, 300194, 300224, 300222, 300168, 300210, 300233, 300207, 300205, 300221, 300203, 300240, 300199, 300200, 300136, 300196, 300178, 300220, 300208, 300126, 300213, 300206, 300219, 300214, 300217, 300215, 300229, 300255, 300257, 300202, 300254, 300223, 300258, 300259, 300218, 300201, 300234, 300237, 300249, 300260, 300238, 300243, 300261, 300231, 300191, 300131, 300236, 300264, 300263, 300193, 300239, 300230, 300235, 300245, 300216, 300247, 300242, 300226, 300270, 300271, 300269, 300174, 300251, 300241, 300232, 300204, 300227, 300228, 300256, 300266, 300281, 300282, 300253, 300244, 300272, 300246, 300145, 300123, 300147, 300297, 300157, 300275, 300284, 300286, 300248, 300273, 300252, 300144, 300146, 300165, 300177, 300173, 300250, 300268, and 300140; UDI-DI 15099590732103.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima28 de octubre de 2026
- FDAROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima28 de octubre de 2026
- FDAIQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345,21 de octubre de 2026
- FDADxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C4951321 de octubre de 2026
- FDANIM Vital Nerve Monitoring systems (all serial numbers manufactured) with NIM Vital System versions: v1.0.0,31 de julio de 2024
- FDAESG-410 Electrosurgical Generator- In conjunction with electrosurgical accessories intended for cutting and31 de julio de 2024
Preguntas sobre este retiro
¿Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2414-2024 el 31 de julio de 2024. Peligro reportado: The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed patient results.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter