Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2416-2024

SlideView VS200 slide viewer,

Publicado el 7 de agosto de 2024 · Inicio del retiro: 22 de julio de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.

Qué hacer

FDA Mandated

Detalles del retiro

Fabricante
EVIDENT SCIENTIFIC INC
Distribuidor
US
Lote
65159 67105 67104 69153 67106 69154
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States