SlideView VS200 slide viewer,
Publicado el 7 de agosto de 2024 · Inicio del retiro: 22 de julio de 2024
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.
Qué hacer
FDA Mandated
Detalles del retiro
- Fabricante
- EVIDENT SCIENTIFIC INC
- Distribuidor
- US
- Lote
- 65159 67105 67104 69153 67106 69154
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States