FDAFDA · Alimentos y medicamentos·N.º oficial Z-2424-2014
LAPAROSCOPY PACK - ( 1) APPLIER MULTI CLIP WITH SHAF GUN L/F ( 1) NEEDLE ULTRA VERES 120MM LIF ( 1) TROCAR ENDOPATH XCEL DILATING TIP (K11LT) LIF ( 1) TROCAR ENDO.XCEL DILATING TIP (KSLT) LIF (1) GOWN IMPERV. XTRA REINFORCED XL T/WRAP AAMI III (1) MAYO STAND COVER REINFORCED LIF (5) SPONGE LAP PREWASH 18" X 18" XRD LIF ( 1) CONTAINER SPECIMEN 40Z. WITH LID & LABEL LIF ( 1) BOWL UTILITY 16oz. LIF (1) SOLUTION SURGICAL DURAPREP 26ML ST. LIF (1) DRAPE LASER CAMERA W/EL 6" X 96" LIF (1 0) GAUZE SPONGE 4" X 4" 16 PLY XRD LIF (4) DRAPE UTILITY WITH TAPE LIF (4) TOWELS ABSORBENT 15" X 20" LIF (1) DRAPE ENDOSCOPY 100" X 123%" X 77" WITH TROUGHS L/F (2) SHEET DRAPE 42" X 57" SMS LIF (1) BLADE SURGICAL #11 STAINLESS STEEL ( 1) TIME OUT BEACON NON WOVEN ST L/F (6) GAUZE SPONGE 2" X 2" 4PL Y LIF ( 1) BLADE SURG. #15 CARBON STEEL ( 1) GOWN XL SMS IMPERVIOUS REINFORCED AAMI Ill LIF (1) TABLE COVER REINFORCED 50" X 90" L/F Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Publicado el 10 de septiembre de 2014 · Inicio del retiro: 20 de mayo de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Customed, Inc
- Distribuidor
- Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
- Lote
- Product code 900-1967, 19 lots: 111123453 112030696 112041219 112041457 112051908 112062470 112083417 112125164 113015692 113025952 113026107 113026233 113047010 113057694 113078292 140211807 140312103 140412626 140412893
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.