Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2506-2026

MEDLINE Medical Procedure Kits labeled as: KIT URO ROBOT CYSTECTOMY, D AND C PACK, MAD D & C, VPH LITHOTOMY MINOR, TRMC- PERI GYN PACK, RICH MINOR LITHOTOMY, etc. (see recall documents for a full list of products)

Publicado el 24 de junio de 2026 · Inicio del retiro: 4 de mayo de 2026

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Qué hacer

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Detalles del retiro

Fabricante
Medline Industries, LP
Distribuidor
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Lote
Medline Kit Number/SKU DYNJ83242B: UDI/DI 10198459157028 (each), 40198459157029 (case), Lot Number 25HBR321; Medline Kit Number/SKU DYNJ83242B: UDI/DI 10198459157028 (each), 40198459157029 (case), Lot Number 25GBD202; Medline Kit Number/SKU DYNJ83242B: UDI/DI 10198459157028 (each), 40198459157029 (case), Lot Number 25FBK813; Medline Kit Number/SKU DYNJ83242B: UDI/DI 10198459157028 (each), 40198459157029 (case), Lot Number 25EBG207; Medline Kit Number/SKU DYNJ61059B: UDI/DI 10195327193454 (each), 40195327193455 (case), Lot Number 23FBK499; Medline Kit Number/SKU DYNJ61059B: UDI/DI 10195327193454 (each), 40195327193455 (case), Lot Number 23EBV071; Medline Kit Number/SKU DYNJ61059B: UDI/DI 10195327193454 (each), 40195327193455 (case), Lot Number 22KBK089; Medline Kit Number/SKU DYNJ61059B: UDI/DI 10195327193454 (each), 40195327193455 (case), Lot Number 22IBV707; Medline Kit Number/SKU DYNJ61059B: UDI/DI 10195327193454 (each), 40195327193455 (case), Lot Number 22HBC310; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 26DMG168; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 26CMD637; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 25KMC849; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 25JMA672; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 25DMC189; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 25CMH655; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 25CMC735; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 24KMA168; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 24IMD969; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 24GMF018; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 24EMA387; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 23LBR964; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 23KBR832; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 23JBE549; Medline Kit Number/SKU DYNJT2135D: UDI/DI 10889942399239 (each), 40889942399230 (case), Lot Number 21LME546; Medline Kit Number/SKU DYNJT2135D: UDI/DI 10889942399239 (each), 40889942399230 (case), Lot Number 21KMG048; Medline Kit Number/SKU DYNJT2135D: UDI/DI 10889942399239 (each), 40889942399230 (case), Lot Number 21JMH240; Medline Kit Number/SKU DYNJT2135D: UDI/DI 10889942399239 (each), 40889942399230 (case), Lot Number 21IMC541; Medline Kit Number/SKU DYNJT2135D: UDI/DI 10889942399239 (each), 40889942399230 (case), Lot Number 21GMD231; Medline Kit Number/SKU DYNJT2135D: UDI/DI 10889942399239 (each), 40889942399230 (case), Lot Number 21DME148; Medline Kit Number/SKU DYNJT2135D: UDI/DI 10889942399239 (each), 40889942399230 (case), Lot Number 21CMC690; Medline Kit Number/SKU DYNJ51750C: UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 25HBB990; Medline Kit Number/SKU DYNJ51750C: UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 25FBO533; Medline Kit Number/SKU DYNJ51750C: UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 25EBA604; Medline Kit Number/SKU DYNJ51750C: UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 25CBB580; Medline Kit Number/SKU DYNJ51750C: UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 25ABE166; Medline Kit Number/SKU DYNJ51750C: UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 24IBR571; Medl
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta