FDAFDA · Alimentos y medicamentos·N.º oficial Z-2507-2024
TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi" Navigation System during surgical procedures in ENT medicine and skull base surgery
Publicado el 14 de agosto de 2024 · Inicio del retiro: 9 de julio de 2024
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Specific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on the system.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Integra LifeSciences Corp.
- Distribuidor
- The devices are only sold in the US. The following states have received the product: AL, AZ CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, PA, SC, SD, TN, TX, VA, VT, WI.
- Lote
- UDI/DI: 10705031245877; Lot Codes: 2310160, 2309110, 2309150, 2308140, 2308160, 2308180.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.