Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2716-2016

NOMAD Pro Packaged X-Ray System, Part Number 08500009; NOMAD Pro Vet Packaged X-Ray System, Part Number 08500011; NOMAD Pro Packaged X-Ray System - Europe, Part Number 08500013; NOMAD Pro X-Ray System, Remote Config., Part Number 08500017; NOMAD Pro Packaged X-Ray System JP, Part Number 08500019

Publicado el 14 de septiembre de 2016 · Inicio del retiro: 18 de julio de 2016

Resumen

FDA registra el retiro Z-2716-2016 de dispositivos médicos de Clase II, publicado el 14 de septiembre de 2016 e iniciado el 18 de julio de 2016. Empresa responsable: Aribex Inc; Estado: Finalizado; Cantidad reportada: 16,871; Lote: Part Number 08500009 - Serial Number Range BX-011008-01 - XB-141219049; Part Number 08500011- Serial Number Range XC -090305-03 - XC-140312-02; Part Number 08500013 - Serial Number Range XB-110411-01 - XB-140616-13; Part Number 08500017 - Serial Number Range XB-120911-04 - XB-141210-50; Part Number 08500019 - Serial Number Range XB-120517-09 - XB-121212-24. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Aribex Inc
Distribuidor
AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA. HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY , Puerto Rico, Afghanistan, Argentina, Australia, Bahamas, Bahrain, Belgium, Bermuda, Chile, China, Columbia, Denmark, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kuwait, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Saudi Arabia, Papua New Guinea, Philippines, Poland, Qatar, Saudi Arabia, Slovenia, South Africa, Sweden, Switzerland, Thailand, Trinidad and Tobago, Turkey, UAE, United Kingdom, Venezuela, Vietnam
Lote
Part Number 08500009 - Serial Number Range BX-011008-01 - XB-141219049; Part Number 08500011- Serial Number Range XC -090305-03 - XC-140312-02; Part Number 08500013 - Serial Number Range XB-110411-01 - XB-140616-13; Part Number 08500017 - Serial Number Range XB-120911-04 - XB-141210-50; Part Number 08500019 - Serial Number Range XB-120517-09 - XB-121212-24
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿NOMAD Pro Packaged X-Ray System, Part Number 08500009; NOMAD Pro Vet Packaged X-Ray System, Part Number 08500011; NOMAD Pro Packaged X-Ray System - Europe, Part Number 08500013; NOMAD Pro X-Ray System, Remote Config., Part Number 08500017; NOMAD Pro Packaged X-Ray System JP, Part Number 08500019 fue retirado del mercado?

Sí. FDA publicó el retiro Z-2716-2016 el 14 de septiembre de 2016. Peligro reportado: Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter