Reusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must occur prior to use. The instruments are intended to be used by qualified physicians to provide access, illumination and visualization of internal structures and for manipulating soft tissue (grasping, cutting, coagulating, dissecting and suturing) in a wide variety of diagnostic and therapeutic laparoscopic/ urologic closed and minimally invasive procedures. It is inserted via a trocar sleeve and is used to prepare, coagulate and cut tissue.
Publicado el 8 de octubre de 2014 · Inicio del retiro: 30 de mayo de 2014
Resumen
FDA registra el retiro Z-2728-2014 de dispositivos médicos de Clase II, publicado el 8 de octubre de 2014 e iniciado el 30 de mayo de 2014. Empresa responsable: Instrumed International, Inc.; Estado: Finalizado; Cantidad reportada: 71 instruments; Lote: 1) Instrumed Item No.: 250-09010; Phoenix Instruments Catalog Code: 190-9523; Lot Number: 070411; 2) Instrumed Item No.: 250-09040; Phoenix Instruments Catalog Code: 190-9527; Lot Numbers: 010211, 040411. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Additional caution statements have been added to the Instructions For Use for monopolar electrodes. 1) To avoid tissue carbonation, the operation voltage of the HF generator must not exceed 650 peak voltage (Vp) for all monopolar electrodes. and 2) For all monopolar electrodes, the useful life of this product is less than or equal to 50 cycles and less than or equal to 2 years.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Instrumed International, Inc.
- Distribuidor
- Distribution US Nationwide in the states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.
- Lote
- 1) Instrumed Item No.: 250-09010; Phoenix Instruments Catalog Code: 190-9523; Lot Number: 070411; 2) Instrumed Item No.: 250-09040; Phoenix Instruments Catalog Code: 190-9527; Lot Numbers: 010211, 040411
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAD'errico perforator drills are bone cutting and drilling instruments that are used without a power source on a17 de diciembre de 2014
- FDAA zipser clamp is an instrument used to compress reduce blood flow to the penis and reduce pain and is sold17 de diciembre de 2014
- FDAMcKenzie enlarging burrs are bone cutting and drilling instruments that are used without a power source on a17 de diciembre de 2014
- FDAGardner-Wells traction tongs are skull tongs for traction used to immobilize a patient with a cervical spine17 de diciembre de 2014
- FDAReusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must8 de octubre de 2014
- FDAReusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must8 de octubre de 2014
Preguntas sobre este retiro
¿Reusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must occur prior to use. The instruments are intended to be used by qualified physicians to provide access, illumination and visualization of internal structures and for manipulating soft tissue (grasping, cutting, coagulating, dissecting and suturing) in a wide variety of diagnostic and therapeutic laparoscopic/ urologic closed and minimally invasive procedures. It is inserted via a trocar sleeve and is used to prepare, coagulate and cut tissue. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2728-2014 el 8 de octubre de 2014. Peligro reportado: Additional caution statements have been added to the Instructions For Use for monopolar electrodes. 1) To avoid tissue carbonation, the operation voltage of the HF generator must not exceed 650 peak voltage (Vp) for all monopolar electrodes. and 2) For all monopolar electrodes, the useful life of this product is less than or equal to 50 cycles and less than or equal to 2 years.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter