13 ft Handpiece Cable, component of the Dermatome Sets Model S and Model S6 and also as a spare part - Product Usage: are intended for use in the cutting of skin grafts for use in plastic surgery and/or burn treatment. These instruments are intended only for professional use in hospitals and similar medical facilities, where the patient will be under the supervision of trained personnel.
Publicado el 23 de septiembre de 2020 · Inicio del retiro: 4 de agosto de 2020
Resumen
FDA registra el retiro Z-2953-2020 de dispositivos médicos de Clase II, publicado el 23 de septiembre de 2020 e iniciado el 4 de agosto de 2020. Empresa responsable: Integra LifeSciences Corp.; Estado: Finalizado; Cantidad reportada: 74 units; Lote: Catalog Number: 353971513 Catalog Description: Handpiece Power Cable Model #: N/A Lot/Serial #: PT-3195. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Integra LifeSciences Corp.
- Distribuidor
- Worldwide distribution - US Nationwide distribution including in the states of MN, TX, CA, UT, WV, FL, IN, GA, AZ, and the countries of Latin America ( CR, MX, TT, EC, PA ), Canada, Europe (FR, AT, BE, ES).
- Lote
- Catalog Number: 353971513 Catalog Description: Handpiece Power Cable Model #: N/A Lot/Serial #: PT-3195
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDABrand Name: MediHoney® Gel Product Name: MediHoney® Wound & Burn Dressing Gel Model/Catalog Number: 3181519 de febrero de 2025
- FDASURG PAT XRAY 1/2X1-1/2 Model/Catalog Number: 801404. The surgical patties and surgical strips are indicated6 de noviembre de 2024
- FDASURG STRP 1-1/2X6 Model/Catalog Number: 801454. The surgical patties and surgical strips are indicated for the6 de noviembre de 2024
- FDASURG PAT XRAY 3/4X3/4 Model/Catalog Number: 801401. The surgical patties and surgical strips are indicated for6 de noviembre de 2024
- FDACarmelized Onion & Gouda Burgers with Creamy Mustard Dressed Corn on the Cob meal kits containing onions23 de septiembre de 2020
- FDAIntegra Padgett Dermatome Set - Product Usage: are intended for use in the cutting of skin grafts for use in23 de septiembre de 2020
Preguntas sobre este retiro
¿13 ft Handpiece Cable, component of the Dermatome Sets Model S and Model S6 and also as a spare part - Product Usage: are intended for use in the cutting of skin grafts for use in plastic surgery and/or burn treatment. These instruments are intended only for professional use in hospitals and similar medical facilities, where the patient will be under the supervision of trained personnel. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2953-2020 el 23 de septiembre de 2020. Peligro reportado: lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter