Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-3058-2017

Medtronic ENT Prgm 9734636 Add Fusion Nav AxiEM, reusable, Rx. This kit includes 9733908 (which includes 9733452 containing 9733449, Straight Suction, 9733450, 70 deg Curved Suction, and 9733451, 90 deg Curved Suction). The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it

Publicado el 30 de agosto de 2017 · Inicio del retiro: 29 de junio de 2017

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Medtronic Navigation, Inc.
Distribuidor
Worldwide Distribution - US (nationwide) and Internationally to Canada.
Lote
Each of the devices within the kit are labeled with information specific to each of the reusable devices within the kit and those devices are what goes into the customer inventory. The outer box where the lot number is printed is discarded by the customer on receipt and the separate devices within the box are what is stored on the shelf. Therefore, no kit lot number is being provided.To be provided
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta