Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.
Publicado el 30 de agosto de 2017 · Inicio del retiro: 23 de mayo de 2017
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Pentax of America Inc
- Distribuidor
- Nationwide
- Lote
- Serial Numbers: G110307, H110482, H110484, H110487, H110489, G110334, G110338, G110404, G110296, G110358, G110340, G110341, G110407, G110409, H110432, G110345, H110450, G110391, H110476, H110443, H110446, G110400, H110483, G110394, G110406, H110411, H110497, G110360, G110363, G110364, G110371, G110366, G110368, G110372, H110434, H110447, G110380, H110455, G110298, H110445, H110449, H110461, G110292, G110346, G110365, H110494, H110495, H110474, G110376, G110319, G110328, H110473, H110413, H110414, H110427, H110453, H110454, H110415, G110312, G110321, H110467, H110471, H110412, G110289, G110303, G110294, H110420, G110299, G110300, H110448, H110452, G110386, G110331, H110451, H110493, G110313, G110314, H110464, H110496, H110433, G110373, G110375, G110317, H110431, H110417, G110291, G110302, G110369, G110378, G110335, G110387, G110390, G110392, G110395, G110311, G110327, G110332, G110382, G110355, G110357, G110287, G110320, H110435, H110481, H110485, H110486, G110354, G110316, G110318, G110362, G110393, H110498, H110456, H110457, H110459, G110337, G110356, H110421, H110429, G110290, G110352, H110470, H110478, G110383, G110384, G110361.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States