System Code Name: 722282 Azurion 7 M20 722281 Azurion 5 M20 722236 Azurion 7 B20 722235 Azurion 7 B12 722234 Azurion 7 M20 722233 Azurion 7 M12 722228 Azurion 5 M20 722227 Azurion 5 M12 722226 Azurion 7 B20 722225 Azurion 7 B12 722224 Azurion 7 M20 722223 Azurion 7 M12 722222 Azurion 3 M15 722221 Azurion 3 M12 722079 Azurion 7 M20 722078 Azurion 7 M12 722068 Azurion 7 B20 722067 Azurion 7 B12 722064 Azurion 3 M15 722063 Azurion 3 M12
Publicado el 25 de septiembre de 2024 · Inicio del retiro: 22 de agosto de 2024
Resumen
FDA registra el retiro Z-3165-2024 de dispositivos médicos de Clase II, publicado el 25 de septiembre de 2024 e iniciado el 22 de agosto de 2024. Empresa responsable: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.; Estado: En curso; Cantidad reportada: 7091 units; Lote: Model Product UDI-DI: 1. 722282 Azurion 7 M20 (01)00884838110571(21); 2. 722281 Azurion 5 M20 (01)00884838110564(21); 3. 722236 Azurion 7 B20 (01)00884838116801(21); 4. 722235 Azurion 7 B12 (01)00884838116788(21); 5. 722234 Azurion 7 M20 (01)00884838116771(21); 6. 722233 Azurion 7 M12 (01)00884838116764(21); 7. 722228 Azurion 5 M20 (01)00884838099234(21); 8. 722227 Azurion 5 M12 (01)00884838099227(21); 9. 722226 Azurion 7 B20 (01)00884838099272(21); 10. 722225 Azurion 7 B12 (01)00884838099265(21); 11. 722224 Azurion 7 M20 (01)00884838099258(21); 12. 722223 Azurion 7 M12 (01)00884838099241(21); 13. 722222 Azurion 3 M15 (01)00884838099210(21); 14. 722221 Azurion 3 M12 (01)00884838099203(21); 15. 722079 Azurion 7 M20 (01)00884838085268(21); 16. 722078 Azurion 7 M12 (01)00884838085251(21); 17. 722068 Azurion 7 B20 (01)00884838085367(21); 18.722067 Azurion 7 B12 (01)00884838085350(21); 19. 722064 Azurion 3 M15 (01)00884838085282(21); 20. 722063 Azurion 3 M12 (01)00884838085275(21) All serial numbers. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Distribuidor
- Nationwide, Canada and Worldwide: Albania Algeria Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Brunei Darussalam Bulgaria Cambodia Cayman Islands Chile China Colombia Costa Rica C®te D'Ivoire Croatia Cuba Curacao Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Ethiopia Fiji Finland France French Polynesia Georgia Germany Ghana Gibraltar Greece Guadeloupe Guinea Honduras Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Jamaica Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Kyrgyzstan Lao People's Democratic Republic Latvia Lebanon Libya Lithuania Macao Macedonia Malaysia Maldives Malta Martinique Mauritania Mauritius Mexico Mongolia Morocco Myanmar Nepal Netherlands New Caledonia New Zealand Nicaragua Norway Oman Pakistan Palestine, State of Panama Papua New Guinea Paraguay Peru Philippines Poland Portugal Qatar R®union Romania Russian Federation Saudi Arabia Senegal Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sudan Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Trinidad and Tobago Tunisia Turkey Turkmenistan Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Venezuela Viet Nam Virgin Islands (U.S.) Yemen
- Lote
- Model Product UDI-DI: 1. 722282 Azurion 7 M20 (01)00884838110571(21); 2. 722281 Azurion 5 M20 (01)00884838110564(21); 3. 722236 Azurion 7 B20 (01)00884838116801(21); 4. 722235 Azurion 7 B12 (01)00884838116788(21); 5. 722234 Azurion 7 M20 (01)00884838116771(21); 6. 722233 Azurion 7 M12 (01)00884838116764(21); 7. 722228 Azurion 5 M20 (01)00884838099234(21); 8. 722227 Azurion 5 M12 (01)00884838099227(21); 9. 722226 Azurion 7 B20 (01)00884838099272(21); 10. 722225 Azurion 7 B12 (01)00884838099265(21); 11. 722224 Azurion 7 M20 (01)00884838099258(21); 12. 722223 Azurion 7 M12 (01)00884838099241(21); 13. 722222 Azurion 3 M15 (01)00884838099210(21); 14. 722221 Azurion 3 M12 (01)00884838099203(21); 15. 722079 Azurion 7 M20 (01)00884838085268(21); 16. 722078 Azurion 7 M12 (01)00884838085251(21); 17. 722068 Azurion 7 B20 (01)00884838085367(21); 18.722067 Azurion 7 B12 (01)00884838085350(21); 19. 722064 Azurion 3 M15 (01)00884838085282(21); 20. 722063 Azurion 3 M12 (01)00884838085275(21) All serial numbers
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Netherlands
Retiros relacionados
- FDAPhilips Azurion 5 M12. Model Number: 722227.4 de noviembre de 2026
- FDAPhilips Azurion 3 M12. Model Numbers: 722063, 722221.4 de noviembre de 2026
- FDAPhilips Azurion 7 M20. Model Numbers: 722079, 722224, 722282.4 de noviembre de 2026
- FDAPhilips Azurion 5 M20. Model Numbers: 722228, 722281.4 de noviembre de 2026
- FDA1. Tech Medical Services Inc. Detachable Endo Retrieval Pouch, REF TM2001, 3" x 6", sterile; 2. Tech Medical25 de septiembre de 2024
- FDASystem Name: 722059 Allura Xper FD20/15 OR Table 722058 Allura Xper FD20/15 722039 Allura Xper FD20/20 OR25 de septiembre de 2024
Preguntas sobre este retiro
¿System Code Name: 722282 Azurion 7 M20 722281 Azurion 5 M20 722236 Azurion 7 B20 722235 Azurion 7 B12 722234 Azurion 7 M20 722233 Azurion 7 M12 722228 Azurion 5 M20 722227 Azurion 5 M12 722226 Azurion 7 B20 722225 Azurion 7 B12 722224 Azurion 7 M20 722223 Azurion 7 M12 722222 Azurion 3 M15 722221 Azurion 3 M12 722079 Azurion 7 M20 722078 Azurion 7 M12 722068 Azurion 7 B20 722067 Azurion 7 B12 722064 Azurion 3 M15 722063 Azurion 3 M12 fue retirado del mercado?
Sí. FDA publicó el retiro Z-3165-2024 el 25 de septiembre de 2024. Peligro reportado: Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter