Ingenia 3.0T and 3.0T CX; Model number (REF): 1. 781342 (3.0T), 2. 781377 (3.0T), 3. 782103 (3.0T), 4. 781271 (3.0T CX); Software Versions: R11, R12 and R13;

Publicado el 7 de octubre de 2026 · Inicio del retiro: 28 de agosto de 2026

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Philips North America
Distribuidor
Worldwide distribution - Nationwide and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Equator. Guinea, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kyrgyzstan, Latvia, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, T¿rkiye, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam.
Lote
1. Model number (REF): 1. 781342 (3.0T); UDI-DI: 00884838055339; Serial Numbers: 71001, 71002, 71008, 71031, 71038, 71051, 71061, 71072, 71084, 71088, 71089, 71095, 71101, 71154, 71163, 71164, 71166, 71184, 71185, 71193, 71210, 71231, 71269, 71271, 71288, 71290, 71297, 71301, 71303, 71309, 71311, 71312, 71322, 71335, 71336, 71340, 71341, 71344, 71355, 71375, 71385, 71393, 71412, 71413, 71420, 71422, 71427, 71438, 71441, 71443, 71445, 71446, 71449, 71450, 71459, 71462, 71463, 71475, 71486, 71487, 71489, 71491, 71513, 71521, 71528, 71556, 71558, 71562, 71589, 71604, 71607, 71612, 71615, 71619, 71620, 71639, 71641, 71646, 71665, 71673, 71688, 71697, 71704, 71710, 71715, 71729, 71730, 71739, 71754, 71756, 71757, 71759, 71813, 71817, 71828, 71831, 71841, 71849, 71850, 71853, 71862, 71885, 71896, 71921, 71923, 71954, 71959, 71960, 71962, 71963, 71964, 71986, 71992, 85006, 85020, 85032, 85038, 85046, 85066, 85068, 85076, 85078, 85085, 85087, 85091, 85092, 85102, 85126, 85148, 85150, 85154, 85191, 85201, 85216, 85245, 85279, 85377; 2. Model Number (REF): 781377 (3.0T); UDI-DI: 00884838009813; Serial Numbers: 42012, 42014, 42015, 42017, 42033, 42036, 42048, 42059, 42088, 42092, 42109, 42125, 42126, 42127, 42146, 42169, 42196, 42198, 42206, 42210, 42218, 42220, 42224, 42228, 42234, 42238, 42247, 42258, 42265, 42271, 42282, 42288, 42293, 42305, 42307, 42321, 42322, 42330, 42342, 42352, 42353, 42359, 42360, 42386, 42389, 42396, 42403, 42409, 42412, 42416, 42425; 3. Model Number (REF): 782103 (3.0T); UDI-DI: 00884838098299; Serial Numbers: 85646, 85652, 85679, 85686, 85708, 85723, 85758, 85759, 85762; 4. Model Number (REF): 781271 (3.0T CX); UDI-DI: 00884838068452; Serial Numbers: 78008, 78009, 78011, 78012, 78017, 78018, 78027, 78028, 78033, 78035, 78036, 78039, 78040, 78042, 78045, 78053, 78060, 78062, 78063, 78067, 78068, 78070, 78095, 78100, 78103, 78112, 78113, 78115, 78122, 78145, 78147, 78148, 78155, 78166, 78167, 78186, 78195, 78201, 78202, 78208, 78215, 78217, 78229, 78277, 78286, 78287;
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Ingenia 3.0T and 3.0T CX; Model number (REF): 1. 781342 (3.0T), 2. 781377 (3.0T), 3. 782103 (3.0T), 4. 781271 (3.0T CX); Software Versions: R11, R12 and R13; fue retirado del mercado?

Sí. FDA publicó el retiro Z-3309-2026 el 7 de octubre de 2026. Peligro reportado: When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter