Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology and Brilliance CT 16 (Air) Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology and Brilliance CT 16 (Air) scanners are whole body Computed Tomography X-ray systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories.
Published October 23, 2013 · Recall initiated: August 30, 2013
Summary
FDA records medical device Class II recall Z-0013-2014, published October 23, 2013 and initiated August 30, 2013. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 258; Lot: System Code #728243, Serial #: 7040, 7083, 7087, 7123, 7124, 7133, 7170, 7178, 7204, 7231, 7273, 7328, 7365, 7370, 7395, 7434, 7467, 7596, 7612, 7775, 7794, 7795, 7840, 7841, 7842, 7843, 7846, 7847, 7848, 7849, 7850, 7851, 7852, 7854, 7855, 7856, 7857, 7858, 7860, 7861, 7863, 7864, 7865, 7866, 7867, 7868, 7871, 7872, 7873, 7875, 7876, 7877, 7881, 7883, 7884, 7885, 7886, 7889, 7890, 7891, 7892, 7893, 7895, 7896, 7898, 7900, 7901, 7902, 7904, 7905, 7907, 7908, 7909, 7910, 7911, 7912, 7914, 7915, 7916, 7918, 7920, 7921, 7923, 7924, 7926, 7927, 7928, 7930, 7931, 7932, 7934, 7935, 7936, 7937, 7938, 7940, 7941, 7942, 7943, 7944 and 7945. System Code #728244, Serial #: 7121, 7271, 7764, 7844, 7845, 7853, 7859, 7862, 7869, 7870, 7874, 7878, 7879, 7880, 7882, 7887, 7888, 7894, 7897, 7899, 7903, 7906, 7913, 7917, 7919 and 7922. System Code #728246, Serial #: 3079, 3090, 3127, 3410, 3478, 3500, 3679, 3721, 5067, 5072, 5073, 5143, 5317, 5439, 5663, 5728, 5741, 5928, 5971, 6028, 6052, 6061, 6067, 6090, 6107, 6133, 6147, 6153, 6156, 6162, 6231, 6491, 6556, 6606, 6607, 6608, 6609, 6610, 6611, 6612, 6613, 6614, 6615, 6616, 6617, 6618, 6619, 6620, 6621, 6622, 6623, 6624, 6625, 6626, 6627, 6628, 6629, 6630, 6631, 6632, 6633, 6634, 6635, 6636, 6637, 6638, 6639, 6640, 6641, 6642, 6643, 6644, 6645, 6646, 6647, 6648, 6649, 6650, 6651, 6652, 6653, 6654, 6655, 6656, 6657, 6658, 6659, 6660, 6661, 6662, 6663, 6664, 6665, 6666, 6667, 6668, 6669, 6670, 6671, 6672, 6673, 6676, 6678, 6679, 6680, 6681, 6682, 6683, 6684, 6685, 6689, 6695, 6696, 6697, 6707, 6710, 6711, 6714, 6721, 30075, 40013, 50026, 50043, 50177, 50181, 50182, 50189, 50195, 50198, 50200 and 60009. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There are artifacts appearing on the Tracker images during threshold-triggered bolus scans with protocols using a Rotation Time of 0.4 seconds.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AR, AZ, CA, CO, CT, GA, HI, IL, IN, KS, MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR,TN, TX, UT, VA, VT and WV., and the countries of Australia, Austria, Bangladesh, Belgium, Brazil, Chile, China, Denmark, Egypt, France, Germany, Honduras, India, Indonesia, Italy, Kazakhstan, Latvia, Lebanon, Libyan Arab Jamahiriya, Mexico, Netherlands, Norway, Palestine, Philippines, Poland, Russia, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom and Vietnam.
- Lot
- System Code #728243, Serial #: 7040, 7083, 7087, 7123, 7124, 7133, 7170, 7178, 7204, 7231, 7273, 7328, 7365, 7370, 7395, 7434, 7467, 7596, 7612, 7775, 7794, 7795, 7840, 7841, 7842, 7843, 7846, 7847, 7848, 7849, 7850, 7851, 7852, 7854, 7855, 7856, 7857, 7858, 7860, 7861, 7863, 7864, 7865, 7866, 7867, 7868, 7871, 7872, 7873, 7875, 7876, 7877, 7881, 7883, 7884, 7885, 7886, 7889, 7890, 7891, 7892, 7893, 7895, 7896, 7898, 7900, 7901, 7902, 7904, 7905, 7907, 7908, 7909, 7910, 7911, 7912, 7914, 7915, 7916, 7918, 7920, 7921, 7923, 7924, 7926, 7927, 7928, 7930, 7931, 7932, 7934, 7935, 7936, 7937, 7938, 7940, 7941, 7942, 7943, 7944 and 7945. System Code #728244, Serial #: 7121, 7271, 7764, 7844, 7845, 7853, 7859, 7862, 7869, 7870, 7874, 7878, 7879, 7880, 7882, 7887, 7888, 7894, 7897, 7899, 7903, 7906, 7913, 7917, 7919 and 7922. System Code #728246, Serial #: 3079, 3090, 3127, 3410, 3478, 3500, 3679, 3721, 5067, 5072, 5073, 5143, 5317, 5439, 5663, 5728, 5741, 5928, 5971, 6028, 6052, 6061, 6067, 6090, 6107, 6133, 6147, 6153, 6156, 6162, 6231, 6491, 6556, 6606, 6607, 6608, 6609, 6610, 6611, 6612, 6613, 6614, 6615, 6616, 6617, 6618, 6619, 6620, 6621, 6622, 6623, 6624, 6625, 6626, 6627, 6628, 6629, 6630, 6631, 6632, 6633, 6634, 6635, 6636, 6637, 6638, 6639, 6640, 6641, 6642, 6643, 6644, 6645, 6646, 6647, 6648, 6649, 6650, 6651, 6652, 6653, 6654, 6655, 6656, 6657, 6658, 6659, 6660, 6661, 6662, 6663, 6664, 6665, 6666, 6667, 6668, 6669, 6670, 6671, 6672, 6673, 6676, 6678, 6679, 6680, 6681, 6682, 6683, 6684, 6685, 6689, 6695, 6696, 6697, 6707, 6710, 6711, 6714, 6721, 30075, 40013, 50026, 50043, 50177, 50181, 50182, 50189, 50195, 50198, 50200 and 60009.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.September 21, 2022
- FDABrilliance 64 Model # 728231, computed tomography x-ray systemAugust 14, 2019
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray systemAugust 14, 2019
- FDABrightView XCT , Model Number 882482July 31, 2019
- FDAGE Healthcare, Precision 500D, Legacy, RFX, and SFX X. This table is intended for use in a diagnostic X-rayOctober 23, 2013
- FDASynthes Midfoot Fusion Bolt 6.5 mm. It is indicated for fracture, fixation, osteotomies, nonunions, andOctober 23, 2013
Questions about this recall
Was Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology and Brilliance CT 16 (Air) Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology and Brilliance CT 16 (Air) scanners are whole body Computed Tomography X-ray systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. recalled?
Yes. FDA published recall Z-0013-2014 on October 23, 2013. Reported hazard: There are artifacts appearing on the Tracker images during threshold-triggered bolus scans with protocols using a Rotation Time of 0.4 seconds.
What should I do if I have this product?
Voluntary: Firm initiated · Letter