Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
Published October 8, 2025 · Recall initiated: August 29, 2025
Summary
FDA records medical device Class II recall Z-0016-2026, published October 8, 2025 and initiated August 29, 2025. Recalling firm: Getinge Disinfection Ab; Status: Ongoing; Reported quantity: 902 units; Lot: Model Number: 88-5. UDI-DI Numbers: 07340153700109, 07340153700116. Software Version: GD 14080 Rev A. All Lot Numbers with electrical heating. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Getinge Disinfection Ab
- Distributor
- Worldwide - US Nationwide distribution in the states of Colorado, Massachusetts, New Mexico, New Hampshire and the countries of Algeria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Austria, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Kenya, Korea, Republic of, Kuwait, Lithuania, Mexico, Netherlands, Norway, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.
- Lot
- Model Number: 88-5. UDI-DI Numbers: 07340153700109, 07340153700116. Software Version: GD 14080 Rev A. All Lot Numbers with electrical heating.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Sweden
Related recalls
- FDAAllura Xper FD20/20 OR Table; Model Numbers: 722039; Associated 510(k)s: K102005, K130638, K130842, K133292,October 8, 2025
- FDANES Reprocessed 2.0mm OTW Turbo Elite Laser Atherectomy Catheter. Model Number: R-420-006. The NES ReprocessedOctober 8, 2025
- FDAAllura Xper FD20/15; Model Numbers: 722058; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979,October 8, 2025
- FDANES Reprocessed 1.7mm RX Turbo Elite Laser Atherectomy Catheter. Model Number: R-417-156. The NES ReprocessedOctober 8, 2025
- FDAAllura Xper FD20/10; Model Numbers: 722029; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979,October 8, 2025
- FDAGETINGE 46-SERIES Medical Washer-Disinfector, Models 46-4 and 46-5, with PACS 300/350 Control Series withDecember 31, 2014
Questions about this recall
Was Getinge 88-Series Washer- Disinfector. Model Number: 88-5. recalled?
Yes. FDA published recall Z-0016-2026 on October 8, 2025. Reported hazard: Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
What should I do if I have this product?
Voluntary: Firm initiated · Letter