Vascular Solutions Twin-Pass RX Dual Access Catheter, model 5210. Sterilized with Ethylene Oxide. The Twin-Pass dual access catheter is a dual lumen catheter designed for use in the coronary and/or peripheral vasculature. The Twin-Pass catheter (Models 5200 and 5230) consists of an over-the-wire (OTW) lumen that runs the length of the catheter and a rapid exchange (RX) delivery lumen on the distal segment. The Twin-Pass catheter has a working length of 135cm and is compatible with various guidewire diameters. Model 5200 also has hydrophilic coating on the distal 18cm of the catheter. The Twin-Pass catheter has white positioning marks located at 95cm (single mark) and 105cm (double marks) from the distal tip, respectively. The Twin-Pass catheter has a marker band located 1mm from the distal tip. Model 5200 contains a second marker band located 11mm from the distal tip to identify the distal end of the OTW lumen.
Published October 26, 2016 · Recall initiated: September 16, 2016
Summary
FDA records medical device Class I recall Z-0054-2017, published October 26, 2016 and initiated September 16, 2016. Recalling firm: Vascular Solutions, Inc.; Status: Terminated; Reported quantity: 11,670 units total; Lot: Model 5210, Lot number: 577762. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Vascular Solutions is recalling Twin-Pass dual access catheters because there is a potential for excess manufacturing material to remain at the tip of the catheter or within the distal portion of the rapid exchange lumen.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Vascular Solutions, Inc.
- Distributor
- Worldwide Distribution: US (nationwide) including Puerto Rico and outside the US to include: Australia, Austria, Belgium, Brazil, Canada, Cyprus, Czech Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Netherlands, Norway, Poland, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.
- Lot
- Model 5210, Lot number: 577762
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADrainer(R) Centesis Catheters: (a) REF 8810, 5F x 9cm (b) REF 8811, 6F x 9cm (c) REF 8812, 5F x 15cm ProductJune 27, 2018
- FDADolphin Inflation Device, REF 0185NA. Rx only, Sterilized using Ethylene Oxide. DOLPHIN is a single-use,June 7, 2017
- FDAGrebSet Micro-Introducer Kit, model 7941, Sterilized with Ethylene Oxide. Product Usage: The GrebSetMarch 1, 2017
- FDAVascular Solutions Twin-Pass .023" Dual Access Catheter, model 5230. Sterilized with Ethylene Oxide. TheOctober 26, 2016
- FDAWilly Rusch GmbH Tracheostomy Tube Set Cannulation of tracheotomised patients, particularly in the case of aOctober 26, 2016
- FDANational Foods Packaging Low Sodium Chicken Base in 50 lb white poly-lined paper bagsOctober 26, 2016
Questions about this recall
Was Vascular Solutions Twin-Pass RX Dual Access Catheter, model 5210. Sterilized with Ethylene Oxide. The Twin-Pass dual access catheter is a dual lumen catheter designed for use in the coronary and/or peripheral vasculature. The Twin-Pass catheter (Models 5200 and 5230) consists of an over-the-wire (OTW) lumen that runs the length of the catheter and a rapid exchange (RX) delivery lumen on the distal segment. The Twin-Pass catheter has a working length of 135cm and is compatible with various guidewire diameters. Model 5200 also has hydrophilic coating on the distal 18cm of the catheter. The Twin-Pass catheter has white positioning marks located at 95cm (single mark) and 105cm (double marks) from the distal tip, respectively. The Twin-Pass catheter has a marker band located 1mm from the distal tip. Model 5200 contains a second marker band located 11mm from the distal tip to identify the distal end of the OTW lumen. recalled?
Yes. FDA published recall Z-0054-2017 on October 26, 2016. Reported hazard: Vascular Solutions is recalling Twin-Pass dual access catheters because there is a potential for excess manufacturing material to remain at the tip of the catheter or within the distal portion of the rapid exchange lumen.
What should I do if I have this product?
Voluntary: Firm initiated · Letter