Dolphin Inflation Device, REF 0185NA. Rx only, Sterilized using Ethylene Oxide. DOLPHIN is a single-use, sterile inflation device used in cardiovascular procedures to pressurize (inflate) and depressurize (deflate) balloon catheters. The device itself consists of two components - the inflation device and large volume syringe (30 cc). The manually operation of the DOLPHIN is achieved by the manipulation of a large handle to drive a piston housed within the body of the device. Careful and controlled inflation is achieved by rotating the handle clockwise. During inflation a unique cam locking mechanism maintains pressure even if the user lets go of the device. Instantaneous deflation, regardless of balloon size, is made possible by the release of the dual locks located on device as well as a large, 30 cc syringe that is sold with each Dolphin Inflation Device. All the while during inflation or deflation pressure is displayed and can be monitored on a large analog gauge mounted on top of the device.
Published June 7, 2017 · Recall initiated: April 12, 2017
Summary
FDA records medical device Class II recall Z-2180-2017, published June 7, 2017 and initiated April 12, 2017. Recalling firm: Vascular Solutions, Inc.; Status: Terminated; Reported quantity: 4,110; Lot: Lot Numbers: 15075668, 15075924, 15115426, 15125175, 16025324, 16035493, 16055261. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The manufacturer of the DOLPHIN inflation device, Perouse Medical, has initiated a recall of the device due to a potential defect in the sterile barrier. Use of affected product could pose a risk of microbial contamination, leading to infection.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Vascular Solutions, Inc.
- Distributor
- US Distribution including PR and to the states of : LA, TN, CA, SC, OR, WA, OK, NV, HI, OH, TX.
- Lot
- Lot Numbers: 15075668, 15075924, 15115426, 15125175, 16025324, 16035493, 16055261.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADrainer(R) Centesis Catheters: (a) REF 8810, 5F x 9cm (b) REF 8811, 6F x 9cm (c) REF 8812, 5F x 15cm ProductJune 27, 2018
- FDAAbbVie PEG, PEG Kit 15FR, FOR ENTERAL USE ONLY, STERILE, Rx Only.June 7, 2017
- FDAThe Acumed Slide-Loc Anatomic Radial Head System is intended for use by surgeons with orthopedic training. TheJune 7, 2017
- FDAGrebSet Micro-Introducer Kit, model 7941, Sterilized with Ethylene Oxide. Product Usage: The GrebSetMarch 1, 2017
- FDAVascular Solutions Twin-Pass .023" Dual Access Catheter, model 5230. Sterilized with Ethylene Oxide. TheOctober 26, 2016
- FDAVascular Solutions Twin-Pass RX Dual Access Catheter, model 5210. Sterilized with Ethylene Oxide. TheOctober 26, 2016
Questions about this recall
Was Dolphin Inflation Device, REF 0185NA. Rx only, Sterilized using Ethylene Oxide. DOLPHIN is a single-use, sterile inflation device used in cardiovascular procedures to pressurize (inflate) and depressurize (deflate) balloon catheters. The device itself consists of two components - the inflation device and large volume syringe (30 cc). The manually operation of the DOLPHIN is achieved by the manipulation of a large handle to drive a piston housed within the body of the device. Careful and controlled inflation is achieved by rotating the handle clockwise. During inflation a unique cam locking mechanism maintains pressure even if the user lets go of the device. Instantaneous deflation, regardless of balloon size, is made possible by the release of the dual locks located on device as well as a large, 30 cc syringe that is sold with each Dolphin Inflation Device. All the while during inflation or deflation pressure is displayed and can be monitored on a large analog gauge mounted on top of the device. recalled?
Yes. FDA published recall Z-2180-2017 on June 7, 2017. Reported hazard: The manufacturer of the DOLPHIN inflation device, Perouse Medical, has initiated a recall of the device due to a potential defect in the sterile barrier. Use of affected product could pose a risk of microbial contamination, leading to infection.
What should I do if I have this product?
Voluntary: Firm initiated · Letter