Leica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue samples

Published October 23, 2024 · Recall initiated: September 4, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
LEICA BIOSYSTEMS NUSSLOCH GMBH
Distributor
US Nationwide distribution.
Lot
UDI/DI 04049188066237, Serial Number 990
Risk classification
Class II
Area (FDA)
device
Country
Germany