Back to search
FDAFDA · Food and drugs·Recall number Z-0088-2025

Leica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c) 14901900700, d) 1491900UVU2; for freezing and sectioning tissue samples

Published October 23, 2024 · Recall initiated: September 4, 2024

Summary

FDA records medical device Class II recall Z-0088-2025, published October 23, 2024 and initiated September 4, 2024. Recalling firm: LEICA BIOSYSTEMS NUSSLOCH GMBH; Status: Ongoing; Reported quantity: 25 units; Lot: a) Material Number 14045227382, UDI/DI 04049188075840, Serial Numbers: 1198121999, 383051996; b) Material Number 14901900100, UDI/DI 04049188003324, Serial Numbers: 703111997, 1018021999, 1538042001, 2058, 629071997, 2994, 1170111999, 3686102007, 3698; c) Material Number 14901900700, UDI/DI 04049188003300, Serial Numbers: 1267032000, 2602062004, 1577052001, 1679092001, 2542042004, 1771, 1885032002, 2682092004, 2376112003, 2774122004, 1114, 1749112001, 1997072002; d) Material Number 1491900UVU2, UDI/DI 04049188002570, Serial Numbers: 3243052006. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
LEICA BIOSYSTEMS NUSSLOCH GMBH
Distributor
US Nationwide distribution.
Lot
a) Material Number 14045227382, UDI/DI 04049188075840, Serial Numbers: 1198121999, 383051996; b) Material Number 14901900100, UDI/DI 04049188003324, Serial Numbers: 703111997, 1018021999, 1538042001, 2058, 629071997, 2994, 1170111999, 3686102007, 3698; c) Material Number 14901900700, UDI/DI 04049188003300, Serial Numbers: 1267032000, 2602062004, 1577052001, 1679092001, 2542042004, 1771, 1885032002, 2682092004, 2376112003, 2774122004, 1114, 1749112001, 1997072002; d) Material Number 1491900UVU2, UDI/DI 04049188002570, Serial Numbers: 3243052006
Risk classification
Class II
Area (FDA)
device
Country
Germany

Related recalls

Questions about this recall

Was Leica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c) 14901900700, d) 1491900UVU2; for freezing and sectioning tissue samples recalled?

Yes. FDA published recall Z-0088-2025 on October 23, 2024. Reported hazard: As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.

What should I do if I have this product?

Voluntary: Firm initiated · Letter