Leica Biosystems Cryostat Model Number CM1950, Material Numbers: a) 14047742467, b) 1491950C1US, c) 1491950C2US, d) 1491950C3US; for freezing and sectioning tissue samples
Published October 23, 2024 · Recall initiated: September 4, 2024
Summary
FDA records medical device Class II recall Z-0089-2025, published October 23, 2024 and initiated September 4, 2024. Recalling firm: LEICA BIOSYSTEMS NUSSLOCH GMBH; Status: Ongoing; Reported quantity: 30 units; Lot: a) Material Number 14047742467, UDI/DI 04049188053992, Serial Numbers: 492102008; b) Material Number 1491950C1US, UDI/DI 04049188002501, Serial Numbers: 432082008, 288032008, 581122008, 435082008, 380, 215012008, 427082008, 218012008, 287032008, 216012008, 291042008; c) Material Number 1491950C2US, UDI/DI 04049188002488, Serial Numbers: 524112008, 451082008, 420, 335, 476, 454, 398, 399072008; d) Material Number 1491950C3US, UDI/DI 04049188002464. Serial Numbers: 314042008, 508, 249032008, 318042008, 429, 328052008, 577122008, 407072008, 391072008, 356062008,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- LEICA BIOSYSTEMS NUSSLOCH GMBH
- Distributor
- US Nationwide distribution.
- Lot
- a) Material Number 14047742467, UDI/DI 04049188053992, Serial Numbers: 492102008; b) Material Number 1491950C1US, UDI/DI 04049188002501, Serial Numbers: 432082008, 288032008, 581122008, 435082008, 380, 215012008, 427082008, 218012008, 287032008, 216012008, 291042008; c) Material Number 1491950C2US, UDI/DI 04049188002488, Serial Numbers: 524112008, 451082008, 420, 335, 476, 454, 398, 399072008; d) Material Number 1491950C3US, UDI/DI 04049188002464. Serial Numbers: 314042008, 508, 249032008, 318042008, 429, 328052008, 577122008, 407072008, 391072008, 356062008,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDALeica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissueOctober 23, 2024
- FDALeica Biosystems Cryostat Model Number CM3050 S, Material Numbers: a) 14047033518, b) 14903050S01, c)October 23, 2024
- FDALeica Biosystems Cryostat Model Number CM1510 S, Material Numbers: a) 491510C002, b) 1491510C003; for freezingOctober 23, 2024
- FDALeica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c)October 23, 2024
- FDALeica Biosystems Cryostat Model Number CM1850, Material Numbers: a) 14047131148, b) 14901850001, c)October 23, 2024
- FDACue Health COVID-19 Test Cartridge REF 2900005 Cue COVID-19 Test Cue COVID-19 Test for Home and Over TheOctober 23, 2024
Questions about this recall
Was Leica Biosystems Cryostat Model Number CM1950, Material Numbers: a) 14047742467, b) 1491950C1US, c) 1491950C2US, d) 1491950C3US; for freezing and sectioning tissue samples recalled?
Yes. FDA published recall Z-0089-2025 on October 23, 2024. Reported hazard: As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
What should I do if I have this product?
Voluntary: Firm initiated · Letter