Soltive Pro SuperPulsed Laser Systems-Intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue Model/Catalog Number: TFL-SLS
Published October 30, 2024 · Recall initiated: August 13, 2024
Summary
FDA records medical device Class II recall Z-0135-2025, published October 30, 2024 and initiated August 13, 2024. Recalling firm: Olympus Corporation of the Americas; Status: Ongoing; Reported quantity: 159 units; Lot: UDI 00821925044135 All Serial Numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Olympus Corporation of the Americas
- Distributor
- Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.
- Lot
- UDI 00821925044135 All Serial Numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOlympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.October 7, 2026
- FDAOlympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.October 7, 2026
- FDAOlympus Tripp-Lite P580-025 25ft DisplayPort Cable, Model Number P580-025, an accessory used in conjunctionOctober 7, 2026
- FDAOlympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector. Model/Catalog Number: CD-B610LA.September 30, 2026
- FDASoltive Premium SuperPulsed Laser - Intended for incision, excision, resection, ablation, coagulation,October 30, 2024
- FDAbk3000, model #2300-11; bk3500, model #2300-41; bk5000, model #2300-61; and bkActive 2300, model #2300-66,October 30, 2024
Questions about this recall
Was Soltive Pro SuperPulsed Laser Systems-Intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue Model/Catalog Number: TFL-SLS recalled?
Yes. FDA published recall Z-0135-2025 on October 30, 2024. Reported hazard: The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.
What should I do if I have this product?
Voluntary: Firm initiated · Letter