REF: P65 ST862 CoLag SCREW, FT, 6.7X62.5 mm
Published October 21, 2026 · Recall initiated: September 9, 2026
Summary
FDA records medical device Class II recall Z-0157-2027, published October 21, 2026 and initiated September 9, 2026. Recalling firm: In2bones USA, LLC; Status: Ongoing; Reported quantity: 29 units; Lot: UDI: 00817906023992/Lot: GM008380; GM007868. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to labeling error, bone screws specified as fully threaded, are actually partially threaded screws.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- In2bones USA, LLC
- Distributor
- US Nationwide distribution in the states of FL, GA, HI, NC, OH, PA, SD, TX, VA.
- Lot
- UDI: 00817906023992/Lot: GM008380; GM007868
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117October 21, 2026
- FDAGeneric Name: GORE VIABIL Short Wire Biliary Endoprosthesis; Model/Catalog Number: VSWVH0808 and VSWVN1006October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116October 21, 2026
- FDAMAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETMOctober 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115October 21, 2026
- FDANeoSpan® SuperElastic Compression Staple w/Instruments Catalog Number and Product Name T50 SN010 NeoSpan®June 17, 2026
Questions about this recall
Was REF: P65 ST862 CoLag SCREW, FT, 6.7X62.5 mm recalled?
Yes. FDA published recall Z-0157-2027 on October 21, 2026. Reported hazard: Due to labeling error, bone screws specified as fully threaded, are actually partially threaded screws.
What should I do if I have this product?
Voluntary: Firm initiated · Letter