Optima NM/CT 640, Models H3100YA, H3100YC. Product Usage: The GE Optima NM/CT 640 system is intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine (NM) system using a variety of scanning modes supported by various acquisition types, and a CT component which is intended specifically for enabling attenuation correction and anatomical localization on SPECT studies
Published November 19, 2014 · Recall initiated: October 14, 2014
Summary
FDA records medical device Class II recall Z-0188-2015, published November 19, 2014 and initiated October 14, 2014. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 29; Lot: 25018 816234D640 26004 815564O640 26007 817250NM640 26010 717782640NM 26012 808485NM640 26013 847872O640 26014 901516GO640 26018 270956NM640 26020 309672NM2 26021 215728OPT640 26025 410543NUC640 26026 928214NMCT 26028 715735OPT1 26002 0910263256 26005 0910263253 22020 030232NU15 26009 DK1054NM03 26001 XM203401 26003 ULUE01NU02 26015 915839NU01 26024 083026800008113 26006 A5639613 26008 A5366308 25023 RTD0782 26030 NM26030 26023 BD003NU01 26017 00162NUC11 26019 00203NUC06 26022 00139NUC08. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide Distribution: US (nationwide) including Guam, Puerto Rico and DC; and countries of: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BELRARUS, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, HUNGARY, GREECE, GUADELOUPE, HONG KONG, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, REPUBLIC OF KOREA, KUWAIT, LATVIA, LEBANON, LITHUANIA, MALAYSIA, MALTA,MARTINIQUE, MEXICO, MOROCCO, NAMIBIA, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN,SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED KINGDOM, UKRAINE, URUGUAY, VENEZUELA, VIET NAAM.
- Lot
- 25018 816234D640 26004 815564O640 26007 817250NM640 26010 717782640NM 26012 808485NM640 26013 847872O640 26014 901516GO640 26018 270956NM640 26020 309672NM2 26021 215728OPT640 26025 410543NUC640 26026 928214NMCT 26028 715735OPT1 26002 0910263256 26005 0910263253 22020 030232NU15 26009 DK1054NM03 26001 XM203401 26003 ULUE01NU02 26015 915839NU01 26024 083026800008113 26006 A5639613 26008 A5366308 25023 RTD0782 26030 NM26030 26023 BD003NU01 26017 00162NUC11 26019 00203NUC06 26022 00139NUC08
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDADiscovery NM 630, dual detector free-geometry integrated nuclear imaging system. Models H3100RC, H3100RD,November 19, 2014
- FDABrivo NM615, is an all-purpose, single detector integrated nuclear imaging system. Models H3100TP, H3100TN,November 19, 2014
Questions about this recall
Was Optima NM/CT 640, Models H3100YA, H3100YC. Product Usage: The GE Optima NM/CT 640 system is intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine (NM) system using a variety of scanning modes supported by various acquisition types, and a CT component which is intended specifically for enabling attenuation correction and anatomical localization on SPECT studies recalled?
Yes. FDA published recall Z-0188-2015 on November 19, 2014. Reported hazard: GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems.
What should I do if I have this product?
Voluntary: Firm initiated · Letter