Brivo NM615, is an all-purpose, single detector integrated nuclear imaging system. Models H3100TP, H3100TN, H3100TT, H3100TR, H3100TW, H3100WZ, H2401MP, H2401MN, H2401MT, H2401MR, H2401MW, H2401NJ. Product Usage: The Brivo NM 615 is an emission computed tomography system intended to detect the location and distribution of gamma ray photon emitting radionuclides in the body and to produce cross-sectional images through computer reconstruction of the data. The Brivo NM 615 system is intended for General Nuclear Medicine imaging procedures using variety of scanning modes supported by various acquisition types. This generic type of device may include signal analysis and display equipment, patient and equipment supports, and accessories
Published November 19, 2014 · Recall initiated: October 14, 2014
Summary
FDA records medical device Class II recall Z-0189-2015, published November 19, 2014 and initiated October 14, 2014. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 22; Lot: Mfg Lot or Serial # System ID 24013 702304NM615 24001 406222LN615 24003 478272NM 24004 405456NM615 24006 806665NM615 24019 030151NU17 24011 NM24011 24014 NM24014 24020 NM24020 24015 NM24015 24016 GON1427120 24002 HC0541NU03 24009 387171NU01 24010 HC3366NU02 24007 083026241010212 00000000116BH4 083026040018413 00000000117BH2 083026020020313 24008 A5635609 24017 NM24017 00000000084BH4 PK1267NM01 24005 ZA2446NM01 24012 ZA1819NM04. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide Distribution: US (nationwide) including Guam, Puerto Rico and DC; and countries of: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BELRARUS, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, HUNGARY, GREECE, GUADELOUPE, HONG KONG, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, REPUBLIC OF KOREA, KUWAIT, LATVIA, LEBANON, LITHUANIA, MALAYSIA, MALTA,MARTINIQUE, MEXICO, MOROCCO, NAMIBIA, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN,SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED KINGDOM, UKRAINE, URUGUAY, VENEZUELA, VIET NAAM.
- Lot
- Mfg Lot or Serial # System ID 24013 702304NM615 24001 406222LN615 24003 478272NM 24004 405456NM615 24006 806665NM615 24019 030151NU17 24011 NM24011 24014 NM24014 24020 NM24020 24015 NM24015 24016 GON1427120 24002 HC0541NU03 24009 387171NU01 24010 HC3366NU02 24007 083026241010212 00000000116BH4 083026040018413 00000000117BH2 083026020020313 24008 A5635609 24017 NM24017 00000000084BH4 PK1267NM01 24005 ZA2446NM01 24012 ZA1819NM04
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDAOptima NM/CT 640, Models H3100YA, H3100YC. Product Usage: The GE Optima NM/CT 640 system is intended forNovember 19, 2014
- FDAMX 16-slice CT Scanner System and MX 16-slice CT Scanner System Phase 11 can be used as a whole body computedNovember 19, 2014
Questions about this recall
Was Brivo NM615, is an all-purpose, single detector integrated nuclear imaging system. Models H3100TP, H3100TN, H3100TT, H3100TR, H3100TW, H3100WZ, H2401MP, H2401MN, H2401MT, H2401MR, H2401MW, H2401NJ. Product Usage: The Brivo NM 615 is an emission computed tomography system intended to detect the location and distribution of gamma ray photon emitting radionuclides in the body and to produce cross-sectional images through computer reconstruction of the data. The Brivo NM 615 system is intended for General Nuclear Medicine imaging procedures using variety of scanning modes supported by various acquisition types. This generic type of device may include signal analysis and display equipment, patient and equipment supports, and accessories recalled?
Yes. FDA published recall Z-0189-2015 on November 19, 2014. Reported hazard: GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems.
What should I do if I have this product?
Voluntary: Firm initiated · Letter