MX 16-slice CT Scanner System and MX 16-slice CT Scanner System Phase 11 can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients. Access Dual CT system can be used as a Head and Whole Body Computed Tomography X-ray System featuring a continuously rotating X-ray tube and detector array with multi-slice capability. The acquired X-ray transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles.
Published November 19, 2014 · Recall initiated: October 9, 2014
Summary
FDA records medical device Class II recall Z-0190-2015, published November 19, 2014 and initiated October 9, 2014. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Ongoing; Reported quantity: 708; Lot: Model Number: 728131, Serial Number(s): EP16E090001,EP16E090002,EP16E090003,EP16E090004,EP16E090006,EP16E090007,EP16E090008, EP16E090009,EP16E090010,EP16E090011,EP16E090012,EP16E090013,EP16E090015,EP16E100001, EP16E100004,EP16E100005,EP16E100006,EP16E100007,EP16E100008,EP16E100010,EP16E100011, EP16E100012,EP16E100013,EP16E100015,EP16E100016,EP16E100017,EP16E100018, EP16E100019,EP16E100020,EP16E100021,EP16E100022,EP16E100023,EP16E100024,EP16E100025, EP16E100026,EP16E100027,EP16E100028,EP16E100029,EP16E100030,EP16E100031,EP16E100032, EP16E100033,EP16E100034,EP16E100035,EP16E100036,EP16E100037,EP16E100038,EP16E100039, EP16E100041,EP16E100043,EP16E100044,EP16E100045,EP16E100046,EP16E100047,EP16E100048, EP16E100049,EP16E100051,EP16E100052,EP16E100055,EP16E100056,EP16E100057,EP16E100058, EP16E100059,EP16E100060,EP16E100061,EP16E100062,EP16E100063,EP16E100067,EP16E100069, EP16E100070,EP16E100071,EP16E100072,EP16E100073,EP16E100074,EP16E100075,EP16E100076, EP16E100077,EP16E100078,EP16E100079,EP16E100080,EP16E100081,EP16E100082,EP16E100083, EP16E100084,EP16E100085,EP16E100086,EP16E100087,EP16E100088,EP16E100090,EP16E100091, EP16E100092,EP16E100093,EP16E100094,EP16E100095,EP16E100096,EP16E100099,EP16E100100, EP16E100101,EP16E100102,EP16E100103,EP16E100104,EP16E100105,EP16E100106,EP16E100107, EP16E100108,EP16E100109,EP16E100110,EP16E100111,EP16E100112,EP16E100113,EP16E110003, EP16E110005,EP16E110006,EP16E110007,EP16E110009,EP16E110011,EP16E110012,EP16E110013, EP16E110014,EP16E110015,EP16E110017,EP16E110018,EP16E110019,EP16E110020,EP16E110021, EP16E110023,EP16E110024,EP16E110025,EP16E110026,EP16E110027,EP16E110028,EP16E110029, EP16E110030,EP16E110031,EP16E110032,EP16E110033,EP16E110034,EP16E110035,EP16E110037, EP16E110038,EP16E110039,EP16E110042,EP16E110043,EP16E110044,EP16E110045,EP16E110046, EP16E110047,EP16E110048,EP16E110049,EP16E110050,EP16E110051,EP16E110052,EP16E110054, EP16E110055,EP16E110056,EP16E110057,EP16E110058,EP16E110060,EP16E110062,EP16E110063, EP16E110064,EP16E110066,EP16E110067,EP16E110069,EP16E110070,EP16E110074,EP16E110075, EP16E110077,EP16E110078,EP16E110083,EP16E110085,EP16E110089,EP16E110090,EP16E110091, EP16E110092,EP16E110093,EP16E110095,EP16E110096,EP16E110097,EP16E110099,EP16E110100, EP16E110101,EP16E110102,EP16E110104,EP16E110105,EP16E110106,EP16E110108,EP16E110109, EP16E110110,EP16E110111,EP16E110115,EP16E110117,EP16E110118,EP16E110119,EP16E120003, EP16E120004,EP16E120005,EP16E120006,EP16E120007,EP16E120008,EP16E120009,EP16E120010, EP16E120011,EP16E120014,EP16E120015,EP16E120016,EP16E120017,EP16E120019,EP16E120021, EP16E120023,EP16E120024,EP16E120025,EP16E120026,EP16E120027,EP16E120029,EP16E120030, EP16E120031,EP16E120032,EP16E120033,EP16E120034,EP16E120035,EP16E120036,EP16E120037, EP16E120039,EP16E120041,EP16E120042,EP16E120043,EP16E120044,EP16E120045,EP16E120046, EP16E120047,EP16E120048,EP16E120050,EP16E120051,EP16E120052,EP16E120054,EP16E120055, EP16E120056,EP16E120058,EP16E120059,EP16E120060,EP16E120061,EP16E120063,EP16E120065, EP16E120067,EP16E120070,EP16E120071,EP16E120072,EP16E120074,EP16E120075,EP16E120076, EP16E120078,EP16E120081,EP16E120083,EP16E120088,EP16E120090,EP16E120091,EP16E120092, EP16E120093,EP16E120095,EP16E120096,EP16E120097,EP16E120098,EP16E120099,EP16E120100, EP16E120101,EP16E120102,EP16E120104,EP16E120105,EP16E120109,EP16E120110,EP16E120111, EP16E120113,EP16E120117,EP16E120118,EP16E120120,EP16E120121,EP16E120122,EP16E120123, EP16E120124,EP16E120125,EP16E120127,EP16E120128,EP16E120129,EP16E120130,EP16E130002, EP16E130003,EP16E130004,EP16E130005,EP16E130006,EP16E130007,EP16E130009,EP16E130010, EP16E130012,EP16E130013,EP16E130014,EP16E130015,EP16E130016,EP16E130017,EP16E130020, EP16E130021,EP16E130022,EP16E130023,EP16E130025,EP16E130026,EP16E130027,EP16E130029, EP16E130030,EP16E130031,EP16E130033,EP16E130034,EP16E130035,EP16E130036,EP16E130037, EP16E130040,EP16E130041,EP16E130042,EP16E130043,EP16E130044,EP16E130045,EP16E130046, EP16E130047,EP16E130048,EP16E130049,EP16E130050,EP16E130051,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
"The orientation of images is displayed incorrectly. When operator chooses gantry on the right side from operation console, and the patient position is HFDL, HFDR, FFDL or FFDR, the patient position picture and text is not consistent. "L" and "R" of scan images are reversed and this is not consistent with the patient actual position. "When a surview which is longe
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- Domestic Distribution: AR, CA, FL, GA, IL, IN, KY, MI, MS, NC, ND, NJ, NV, NY, OH, PA, PR, SC, TN, and VA. Foreign distribution: Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Chile, China, Colombia, Congo, Ecuador, Egypt, France, Germany, Ghana, Greece, Gutamala, Hungray, India, Indonesia, Ireland, Italy, Japan, Jordan, Latvia, Lybyan Arab JHamahiri, Lithuania, Malawi, Mexico, Myanmar, Namibia, Nepal, Netherlands, Oman, Palestine, Paraguay, Philippines, Poland, Portugal, Romania,Russia, Rwanda, Saudi Arabia, Senegal, South Agrica, South ZKorea, Spain, Sri Lanka, Switzerland, Tunisia, Turkey, Uganda, UAE, Venezuela, Vietnam, Zambia.
- Lot
- Model Number: 728131, Serial Number(s): EP16E090001,EP16E090002,EP16E090003,EP16E090004,EP16E090006,EP16E090007,EP16E090008, EP16E090009,EP16E090010,EP16E090011,EP16E090012,EP16E090013,EP16E090015,EP16E100001, EP16E100004,EP16E100005,EP16E100006,EP16E100007,EP16E100008,EP16E100010,EP16E100011, EP16E100012,EP16E100013,EP16E100015,EP16E100016,EP16E100017,EP16E100018, EP16E100019,EP16E100020,EP16E100021,EP16E100022,EP16E100023,EP16E100024,EP16E100025, EP16E100026,EP16E100027,EP16E100028,EP16E100029,EP16E100030,EP16E100031,EP16E100032, EP16E100033,EP16E100034,EP16E100035,EP16E100036,EP16E100037,EP16E100038,EP16E100039, EP16E100041,EP16E100043,EP16E100044,EP16E100045,EP16E100046,EP16E100047,EP16E100048, EP16E100049,EP16E100051,EP16E100052,EP16E100055,EP16E100056,EP16E100057,EP16E100058, EP16E100059,EP16E100060,EP16E100061,EP16E100062,EP16E100063,EP16E100067,EP16E100069, EP16E100070,EP16E100071,EP16E100072,EP16E100073,EP16E100074,EP16E100075,EP16E100076, EP16E100077,EP16E100078,EP16E100079,EP16E100080,EP16E100081,EP16E100082,EP16E100083, EP16E100084,EP16E100085,EP16E100086,EP16E100087,EP16E100088,EP16E100090,EP16E100091, EP16E100092,EP16E100093,EP16E100094,EP16E100095,EP16E100096,EP16E100099,EP16E100100, EP16E100101,EP16E100102,EP16E100103,EP16E100104,EP16E100105,EP16E100106,EP16E100107, EP16E100108,EP16E100109,EP16E100110,EP16E100111,EP16E100112,EP16E100113,EP16E110003, EP16E110005,EP16E110006,EP16E110007,EP16E110009,EP16E110011,EP16E110012,EP16E110013, EP16E110014,EP16E110015,EP16E110017,EP16E110018,EP16E110019,EP16E110020,EP16E110021, EP16E110023,EP16E110024,EP16E110025,EP16E110026,EP16E110027,EP16E110028,EP16E110029, EP16E110030,EP16E110031,EP16E110032,EP16E110033,EP16E110034,EP16E110035,EP16E110037, EP16E110038,EP16E110039,EP16E110042,EP16E110043,EP16E110044,EP16E110045,EP16E110046, EP16E110047,EP16E110048,EP16E110049,EP16E110050,EP16E110051,EP16E110052,EP16E110054, EP16E110055,EP16E110056,EP16E110057,EP16E110058,EP16E110060,EP16E110062,EP16E110063, EP16E110064,EP16E110066,EP16E110067,EP16E110069,EP16E110070,EP16E110074,EP16E110075, EP16E110077,EP16E110078,EP16E110083,EP16E110085,EP16E110089,EP16E110090,EP16E110091, EP16E110092,EP16E110093,EP16E110095,EP16E110096,EP16E110097,EP16E110099,EP16E110100, EP16E110101,EP16E110102,EP16E110104,EP16E110105,EP16E110106,EP16E110108,EP16E110109, EP16E110110,EP16E110111,EP16E110115,EP16E110117,EP16E110118,EP16E110119,EP16E120003, EP16E120004,EP16E120005,EP16E120006,EP16E120007,EP16E120008,EP16E120009,EP16E120010, EP16E120011,EP16E120014,EP16E120015,EP16E120016,EP16E120017,EP16E120019,EP16E120021, EP16E120023,EP16E120024,EP16E120025,EP16E120026,EP16E120027,EP16E120029,EP16E120030, EP16E120031,EP16E120032,EP16E120033,EP16E120034,EP16E120035,EP16E120036,EP16E120037, EP16E120039,EP16E120041,EP16E120042,EP16E120043,EP16E120044,EP16E120045,EP16E120046, EP16E120047,EP16E120048,EP16E120050,EP16E120051,EP16E120052,EP16E120054,EP16E120055, EP16E120056,EP16E120058,EP16E120059,EP16E120060,EP16E120061,EP16E120063,EP16E120065, EP16E120067,EP16E120070,EP16E120071,EP16E120072,EP16E120074,EP16E120075,EP16E120076, EP16E120078,EP16E120081,EP16E120083,EP16E120088,EP16E120090,EP16E120091,EP16E120092, EP16E120093,EP16E120095,EP16E120096,EP16E120097,EP16E120098,EP16E120099,EP16E120100, EP16E120101,EP16E120102,EP16E120104,EP16E120105,EP16E120109,EP16E120110,EP16E120111, EP16E120113,EP16E120117,EP16E120118,EP16E120120,EP16E120121,EP16E120122,EP16E120123, EP16E120124,EP16E120125,EP16E120127,EP16E120128,EP16E120129,EP16E120130,EP16E130002, EP16E130003,EP16E130004,EP16E130005,EP16E130006,EP16E130007,EP16E130009,EP16E130010, EP16E130012,EP16E130013,EP16E130014,EP16E130015,EP16E130016,EP16E130017,EP16E130020, EP16E130021,EP16E130022,EP16E130023,EP16E130025,EP16E130026,EP16E130027,EP16E130029, EP16E130030,EP16E130031,EP16E130033,EP16E130034,EP16E130035,EP16E130036,EP16E130037, EP16E130040,EP16E130041,EP16E130042,EP16E130043,EP16E130044,EP16E130045,EP16E130046, EP16E130047,EP16E130048,EP16E130049,EP16E130050,EP16E130051,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.September 21, 2022
- FDABrilliance 64 Model # 728231, computed tomography x-ray systemAugust 14, 2019
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray systemAugust 14, 2019
- FDABrightView XCT , Model Number 882482July 31, 2019
- FDABrivo NM615, is an all-purpose, single detector integrated nuclear imaging system. Models H3100TP, H3100TN,November 19, 2014
- FDABard® PerFix Light Plug The Bard® PerFix Light Plug is indicated for reinforcement if soft tissue, whereNovember 19, 2014
Questions about this recall
Was MX 16-slice CT Scanner System and MX 16-slice CT Scanner System Phase 11 can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients. Access Dual CT system can be used as a Head and Whole Body Computed Tomography X-ray System featuring a continuously rotating X-ray tube and detector array with multi-slice capability. The acquired X-ray transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. recalled?
Yes. FDA published recall Z-0190-2015 on November 19, 2014. Reported hazard: "The orientation of images is displayed incorrectly. When operator chooses gantry on the right side from operation console, and the patient position is HFDL, HFDR, FFDL or FFDR, the patient position picture and text is not consistent. "L" and "R" of scan images are reversed and this is not consistent with the patient actual position. "When a surview which is longe
What should I do if I have this product?
Voluntary: Firm initiated · Letter