Back to search
FDAFDA · Food and drugs·Recall number Z-0195-2015

Series OEC 9800. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient,during diagnostic, surgical and interventional procedures.

Published November 19, 2014 · Recall initiated: October 3, 2014

Summary

FDA records medical device Class II recall Z-0195-2015, published November 19, 2014 and initiated October 3, 2014. Recalling firm: GE OEC Medical Systems, Inc; Status: Terminated; Reported quantity: 311; Lot: Serial No: 82-0013,82-0045,82-0061,82-0071,82-0085,82-0122,82-0133,82-0190,82-0246,82-0357,82-0528,82-0533,82-0597,82-0626,82-0699,82-0702,82-0707,82-0718,82-0805,82-0818,82-0854,82-0855,82-0946,82-1040,82-1052,82-1062,82-1106,82-1133,82-1199,82-1292,82-1324,82-1348,82-1411,82-1453,82-1460,82-1526,82-1542,82-1608,82-1651,82-1658,82-1679,82-1709,82-1731,82-1773,82-1822,82-1867,82-1882,82-1905,82-1914,82-1979,82-2017,82-2018,82-2024,82-2030,82-2036,82-2049,82-2165,82-2169,82-2206,82-2225,82-2241,82-2279,82-2356,82-2388,82-2398,82-2454,82-2455,82-2459,82-2474,82-2535,82-2565,82-2603,82-2650,82-2655,82-2735,82-2751,82-2761,82-2767,82-2783,82-2851,82-2855,82-2909,82-2930,82-2945,82-2975,82-3003,82-3094,82-3096,82-3169,82-3210,82-3226,82-3232,82-3241,82-3319,82-3325,82-3377,82-3389,82-3395,82-3406,82-3424,82-3430,82-3456,82-3459,82-3500,82-3523,82-3534,82-3539,82-3569,82-3571,82-3629,82-3713,82-3748,82-3774,82-3826,82-3837,82-3851,82-3895,82-7048-MH,82-7066-MH,82-7103-MH,89-0028,89-0034,89-0037,89-0072,89-0105,89-0113,89-0135-RC,89-0184,89-0192,89-0308,89-0311,89-0316,89-0318,89-0388,89-0396,89-0427,89-0517,89-0586,89-0600,89-0623,89-0624,89-0642,89-0643,89-0743,89-0825,89-0827,89-0840,89-0843,89-0846,89-0866,89-0881,89-0894,89-0930,89-0947,89-0995,89-1023,89-1145,89-1157,89-1202,89-1204,89-1209,89-1260,89-1308,89-1327,89-1375,89-1392,89-1408,89-1411,89-1544,89-1652,89-1671,89-1698,89-1751,89-1804,89-1849,89-1868,89-1956,89-1964,89-1984,89-1986,89-2031,89-2050,89-2083,89-2085,89-2094,89-2118,89-2142,89-2170,89-2199,89-2244,89-2250,89-2336,89-2354,89-2459,89-2550,89-2555,89-2598,89-2623,89-2674,89-2682,89-2690,89-2695,89-2767,89-2777,89-2850,89-2862,89-2900,89-3031,89-3032,89-3124-N,89-3165,89-3171,89-3222,89-3252,89-3254,89-3311,89-3318,89-3346,89-3359,89-3367,89-3368,89-3419,89-3438,89-3531,8S-0014-C,8S-0026,8S-0029,8S-0058,8S-0059,8S-0070,8S-0113,8S-0162,8S-0224,8S-0232,8S-0389,8S-0467,8S-0511,8S-0529,8S-0531,8S-0549,8S-0605,8S-0755,8S-0766,8S-0782,8S-0858,8S-0876,8S-0933-C,8S-0997,8S-1000,8S-1015,8S-1147,8S-1165,8S-1166,8S-1192,8S-1427,8S-1447,8S-1509-C,8S-1565,8S-1585,8S-1897,8S-1907,8S-2008-N,8S-2053,8S-2062,8S-2092,8S-2103,8S-2107,8S-2232,8S-2234,8S-2268,8S-2310,8S-2385,8S-2408,8S-2432,8S-2445,8S-2471-C,8S-2491,8S-2530,8S-2543,8S-2550,8S-2554,8S-2564,8S-2618,8S-2650,8S-2651,8S-2671,8S-2675,8S-2695,8S-2701,8S-2734,8S-2746,8S-2770,8S-2792,8S-2798,8S-2819,8S-2835,8S-2861,8S-2862,8S-2914,8S-2922,8S-2984,8S-3004,8S-3048,8S-3055,8S-3088,8S-3143,8S-3162-H,8S-3219,8S-3272-H,8S-3289-H,8S-7102-MH,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE OEC Medical Systems, Inc
Distributor
Worldwide Distribution.
Lot
Serial No: 82-0013,82-0045,82-0061,82-0071,82-0085,82-0122,82-0133,82-0190,82-0246,82-0357,82-0528,82-0533,82-0597,82-0626,82-0699,82-0702,82-0707,82-0718,82-0805,82-0818,82-0854,82-0855,82-0946,82-1040,82-1052,82-1062,82-1106,82-1133,82-1199,82-1292,82-1324,82-1348,82-1411,82-1453,82-1460,82-1526,82-1542,82-1608,82-1651,82-1658,82-1679,82-1709,82-1731,82-1773,82-1822,82-1867,82-1882,82-1905,82-1914,82-1979,82-2017,82-2018,82-2024,82-2030,82-2036,82-2049,82-2165,82-2169,82-2206,82-2225,82-2241,82-2279,82-2356,82-2388,82-2398,82-2454,82-2455,82-2459,82-2474,82-2535,82-2565,82-2603,82-2650,82-2655,82-2735,82-2751,82-2761,82-2767,82-2783,82-2851,82-2855,82-2909,82-2930,82-2945,82-2975,82-3003,82-3094,82-3096,82-3169,82-3210,82-3226,82-3232,82-3241,82-3319,82-3325,82-3377,82-3389,82-3395,82-3406,82-3424,82-3430,82-3456,82-3459,82-3500,82-3523,82-3534,82-3539,82-3569,82-3571,82-3629,82-3713,82-3748,82-3774,82-3826,82-3837,82-3851,82-3895,82-7048-MH,82-7066-MH,82-7103-MH,89-0028,89-0034,89-0037,89-0072,89-0105,89-0113,89-0135-RC,89-0184,89-0192,89-0308,89-0311,89-0316,89-0318,89-0388,89-0396,89-0427,89-0517,89-0586,89-0600,89-0623,89-0624,89-0642,89-0643,89-0743,89-0825,89-0827,89-0840,89-0843,89-0846,89-0866,89-0881,89-0894,89-0930,89-0947,89-0995,89-1023,89-1145,89-1157,89-1202,89-1204,89-1209,89-1260,89-1308,89-1327,89-1375,89-1392,89-1408,89-1411,89-1544,89-1652,89-1671,89-1698,89-1751,89-1804,89-1849,89-1868,89-1956,89-1964,89-1984,89-1986,89-2031,89-2050,89-2083,89-2085,89-2094,89-2118,89-2142,89-2170,89-2199,89-2244,89-2250,89-2336,89-2354,89-2459,89-2550,89-2555,89-2598,89-2623,89-2674,89-2682,89-2690,89-2695,89-2767,89-2777,89-2850,89-2862,89-2900,89-3031,89-3032,89-3124-N,89-3165,89-3171,89-3222,89-3252,89-3254,89-3311,89-3318,89-3346,89-3359,89-3367,89-3368,89-3419,89-3438,89-3531,8S-0014-C,8S-0026,8S-0029,8S-0058,8S-0059,8S-0070,8S-0113,8S-0162,8S-0224,8S-0232,8S-0389,8S-0467,8S-0511,8S-0529,8S-0531,8S-0549,8S-0605,8S-0755,8S-0766,8S-0782,8S-0858,8S-0876,8S-0933-C,8S-0997,8S-1000,8S-1015,8S-1147,8S-1165,8S-1166,8S-1192,8S-1427,8S-1447,8S-1509-C,8S-1565,8S-1585,8S-1897,8S-1907,8S-2008-N,8S-2053,8S-2062,8S-2092,8S-2103,8S-2107,8S-2232,8S-2234,8S-2268,8S-2310,8S-2385,8S-2408,8S-2432,8S-2445,8S-2471-C,8S-2491,8S-2530,8S-2543,8S-2550,8S-2554,8S-2564,8S-2618,8S-2650,8S-2651,8S-2671,8S-2675,8S-2695,8S-2701,8S-2734,8S-2746,8S-2770,8S-2792,8S-2798,8S-2819,8S-2835,8S-2861,8S-2862,8S-2914,8S-2922,8S-2984,8S-3004,8S-3048,8S-3055,8S-3088,8S-3143,8S-3162-H,8S-3219,8S-3272-H,8S-3289-H,8S-7102-MH,
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Series OEC 9800. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient,during diagnostic, surgical and interventional procedures. recalled?

Yes. FDA published recall Z-0195-2015 on November 19, 2014. Reported hazard: GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.

What should I do if I have this product?

Voluntary: Firm initiated · Letter