FDAFDA · Food and drugs·Recall number Z-0347-2027

Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60),

Published November 11, 2026 · Recall initiated: September 30, 2025

Summary

FDA records medical device Class II recall Z-0347-2027, published November 11, 2026 and initiated September 30, 2025. Recalling firm: SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION; Status: Ongoing; Reported quantity: 118 unts.; Lot: Serial number/Shimadzu Medical Systems #/UDI: M1F0D3BF4001/ SMS20250167/ (01)04540217067145(11)250; M1F0CC4EB001/ SMS20250098/ (01)04540217067022(11)241; M1F0D27EA001/ SMS20250001/ (01)04540217067138(11)241; 41F0F43E7001/ SMS20250034/ (01)04540217068951(11)240; M1F0D3BE7001/ SMS20250248/ (01)04540217067145(11)240; M1F0D3BE6001/ SMS20240196/ (01)04540217067145(11)240; 41F0F40E5001/ N/A/ (01)04540217068937(11)240; 41F0F41E5001/ N/A/ (01)04540217068944(11)240; 41F0F43E5001/ SMS20240236/ (01)04540217068951(11)240; M1F0CD7E4001/ SMS20240162/ (01)04540217067046(11)240; 41F0F41E1001/ SMS20240062/ (01)04540217068944(11)240; 41F0F43DB002/ SMS20240101/ (01)04540217068951(11)231; 41F0F43DB001/ SMS20240100/ (01)04540217068951(11)231; M1F0CD7D4001/ SMS20230178/ (01)04540217067046(11)230. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION
Distributor
US Nationwide distribution in the states of SC, TX, FL, TN, CA, FL, NY.
Lot
Serial number/Shimadzu Medical Systems #/UDI: M1F0D3BF4001/ SMS20250167/ (01)04540217067145(11)250; M1F0CC4EB001/ SMS20250098/ (01)04540217067022(11)241; M1F0D27EA001/ SMS20250001/ (01)04540217067138(11)241; 41F0F43E7001/ SMS20250034/ (01)04540217068951(11)240; M1F0D3BE7001/ SMS20250248/ (01)04540217067145(11)240; M1F0D3BE6001/ SMS20240196/ (01)04540217067145(11)240; 41F0F40E5001/ N/A/ (01)04540217068937(11)240; 41F0F41E5001/ N/A/ (01)04540217068944(11)240; 41F0F43E5001/ SMS20240236/ (01)04540217068951(11)240; M1F0CD7E4001/ SMS20240162/ (01)04540217067046(11)240; 41F0F41E1001/ SMS20240062/ (01)04540217068944(11)240; 41F0F43DB002/ SMS20240101/ (01)04540217068951(11)231; 41F0F43DB001/ SMS20240100/ (01)04540217068951(11)231; M1F0CD7D4001/ SMS20230178/ (01)04540217067046(11)230
Risk classification
Class II
Area (FDA)
device
Country
Japan

Related recalls

Questions about this recall

Was Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60), recalled?

Yes. FDA published recall Z-0347-2027 on November 11, 2026. Reported hazard: The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail