Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner
Published November 11, 2026 · Recall initiated: October 1, 2026
Summary
FDA records medical device Class II recall Z-0348-2027, published November 11, 2026 and initiated October 1, 2026. Recalling firm: restor3d Inc.; Status: Ongoing; Reported quantity: 240 units; Lot: All Lot codes/ UDIs:Product Code: 3500-140C3 UDI/GTIN:00840097513492; Product Code: 3500-140U3 UDI/GTIN:00840097513461; Product Code: 3500-140C1 UDI/GTIN:00840097513430; Product Code: 3500-140U1 UDI/GTIN:00840097513409. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- restor3d Inc.
- Distributor
- US Nationwide distribution in the states of Texas, Massachusetts, Oklahoma, Florida, Colorado, North Carolina.
- Lot
- All Lot codes/ UDIs:Product Code: 3500-140C3 UDI/GTIN:00840097513492; Product Code: 3500-140U3 UDI/GTIN:00840097513461; Product Code: 3500-140C1 UDI/GTIN:00840097513430; Product Code: 3500-140U1 UDI/GTIN:00840097513409
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATrinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-RayNovember 11, 2026
- FDACochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DINovember 11, 2026
- FDAGeneric Name: Perma Sharp Sutures; Generic Name: Polypropylene Nonabsorbable Surgical Suture; Model/CatalogNovember 11, 2026
- FDAGeneric Name: restor3d MTP Implant Generic Name: MTP Implant Model/Catalog Number: 1103-161620 SoftwareOctober 28, 2026
- FDAKinos Axiom Total Ankle System with Axiom PSR, Flat-cut (Cut Guide), REF: 5831-9006September 23, 2026
- FDABrand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid BaseplateAugust 26, 2026
Questions about this recall
Was Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner recalled?
Yes. FDA published recall Z-0348-2027 on November 11, 2026. Reported hazard: (Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail