FDAFDA · Food and drugs·Recall number Z-0357-2027

intelerad IntelePACS software

Published November 11, 2026 · Recall initiated: September 10, 2026

Summary

FDA records medical device Class II recall Z-0357-2027, published November 11, 2026 and initiated September 10, 2026. Recalling firm: INTELERAD MEDICAL SYSTEMS INCORPORATED; Status: Ongoing; Reported quantity: 560 software platforms; Lot: HIBCC:B228INTELEPACS0/Affected versions: PACS-4-3-4-R20 PACS-4-9-1-R80 PACS-4-12-1-R81 through R90 PACS-4-14-1-R46 through R137 PACS-4-16-1-R50 through R137 PACS-4-17-1-R23 through R117 PACS-4-18-1-R31 through R91 PACS-5-1-1-R17 through R102 PACS-5-2-1-R36 through R93 PACS-5-3-1-R26 through R128 PACS-5-4-1-R15 through R101 PACS-5-5-1-R7 through R89 PACS-5-6-1-R12 through R81 PACS-5-7-1-R2 through R61 PACS-5-8-1-R2 through R42 PACS-5-9-1-R8 through R19. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A cybersecurity vulnerability has been identified in the software that could allow a malicious user to take over an administrator session, delete or modify image or study data, or lock legitimate users out of the system. Additionally, a software issue has been identified that may result in the unintended deletion of image data from an external archive integration.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
INTELERAD MEDICAL SYSTEMS INCORPORATED
Distributor
US: AL FL CO MI CA TX CT AZ NJ IL NC SC MO NH WA KY WY DC NV IN PA AK OR AR KS TN MN MD IA WI HI NY OH VA WV NM GA LA NE UT WA OUS: Australia , Canada, Ireland, New Zealand, South Africa, United Kingdom
Lot
HIBCC:B228INTELEPACS0/Affected versions: PACS-4-3-4-R20 PACS-4-9-1-R80 PACS-4-12-1-R81 through R90 PACS-4-14-1-R46 through R137 PACS-4-16-1-R50 through R137 PACS-4-17-1-R23 through R117 PACS-4-18-1-R31 through R91 PACS-5-1-1-R17 through R102 PACS-5-2-1-R36 through R93 PACS-5-3-1-R26 through R128 PACS-5-4-1-R15 through R101 PACS-5-5-1-R7 through R89 PACS-5-6-1-R12 through R81 PACS-5-7-1-R2 through R61 PACS-5-8-1-R2 through R42 PACS-5-9-1-R8 through R19
Risk classification
Class II
Area (FDA)
device
Country
Canada

Related recalls

Questions about this recall

Was intelerad IntelePACS software recalled?

Yes. FDA published recall Z-0357-2027 on November 11, 2026. Reported hazard: A cybersecurity vulnerability has been identified in the software that could allow a malicious user to take over an administrator session, delete or modify image or study data, or lock legitimate users out of the system. Additionally, a software issue has been identified that may result in the unintended deletion of image data from an external archive integration.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail