ORTHOFIX Catalogue Number: ref 1-1100701,DRILL BIT D.4,8 MM L.280 MM TIN COATED - QUICK CONNECT, RX ONLY, Non Sterile, UDI: (01)18033509856099
Published November 20, 2019 · Recall initiated: September 27, 2018
Summary
FDA records medical device Class II recall Z-0383-2020, published November 20, 2019 and initiated September 27, 2018. Recalling firm: Orthofix Srl; Status: Terminated; Reported quantity: N/A; Lot: Lot # B1199835, 289815, 289815, 273344. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Orthofix Srl
- Distributor
- US: AZ, CA,FL,GA,HI,IA,IL,IN,KY,LA,MD,MI,MO,MS,NC,NY,OH,OK,OR,PA,SC,TN,TX,UT,VA,WA, WI, WV, Puerto Rico OUS: Kuwait, Australia, Saudi Arabia, United Arab Emirates, Ireland, Japan, Spain, Poland, Switzerland, Singapore, Peru, Brazil, Germany, Great Britain, France, Italy, Australia, South Africa, Belgium, Chile, The Netherlands, Lithuania, Sri Lanka, Denmark, Israel
- Lot
- Lot # B1199835, 289815, 289815, 273344
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Italy
Related recalls
- FDAREF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, SterileJune 25, 2025
- FDAORTHOFIX Catalogue Number: ref 1100701, drill bit diameter 4.8 MM LENGTH 280MM, RX ONLY, Non Sterile, UDI:November 20, 2019
- FDAORTHOFIX Catalogue Number: ref: 99-92501, PREFIX II DIAPHYSEAL COMPLETE KIT STERILE SHORT SCREWS, RX ONLY,November 20, 2019
- FDARANDOX NEFA (Non-Esterified Fatty Acids) FA115 Lot 485343, GTIN: 5055273203066.November 20, 2019
- FDAVisual-ICE Cryoablation System, FPRCH6000. The cryoablation system is a mobile console system intended forNovember 20, 2019
- FDATL-HEX TruLok Hexapod System, Double Row Footplates, TL-HEX Sterile, All Sizes: 120 MM, 140 MM, 160 MM, 180June 13, 2018
Questions about this recall
Was ORTHOFIX Catalogue Number: ref 1-1100701,DRILL BIT D.4,8 MM L.280 MM TIN COATED - QUICK CONNECT, RX ONLY, Non Sterile, UDI: (01)18033509856099 recalled?
Yes. FDA published recall Z-0383-2020 on November 20, 2019. Reported hazard: As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
What should I do if I have this product?
Voluntary: Firm initiated · Letter